90-0322-AAA
OHG Sub-Unit Installation Instructions
20th Nov 2012
© 2012 Owlstone Ltd
Proprietary and Confidential
Page 8 of 14
Section 6 -
Installing a Quick Connect in GEN-SYS
Warning: Swagelok fittings will leak if fitted incorrectly. Owlstone will not take any
responsibility for incorrect fitting of Swagelok fittings by the end user.
The user must undertake a risk assessment before following this procedure. The user is
responsible for assessing chemical safety, COSHH and other risk implications.
If the user requires the GEN-SYS system to be leak tight it must be tested after assembly.
Please contact Owlstone for technical support regarding leak testing.
1) Prepare materials
A Swagelok Quick Connect is supplied with your OHG (Owlstone p/n 50-0081).
Two
7/16” and a 5/8” spanner are required for this procedure.
Figure 10: Quick connect
2) Remove Swagelok cap, fit Quick Connect
Remove the
Swagelok cap using the two 7/16” spanners before installing the quick connect. Swagelok
fittings must be tightened by hand before turning an additional 1x half turn using a 7/16
” spanner.
Do not over-tighten.
Figure 11: Remove the cap
Figure 12: Install the quick connect