7.4
Baking O-rings
Note: o-rings must NEVER be greased or coated as this will add
contaminants or could change the sealing properties of the o-ring.
Note: The o-rings supplied with this kit have been pre-baked.
Any o-rings
not
supplied by Owlstone Medical should also be baked
to reduce the contaminants they hold.
This is done by arranging them on a clean, stainless steel try, slightly
overlapping to decrease the area touching the tray.
Bake in a clean oven (preferably only used for this purpose) at 180C
for 12-16 hours
, preferably with carbon scrubbed or “zero air” flowing
through the oven to remove the contaminants that evolve from the o-
rings.
7.5
Cleaning the 01-0603 Travel End Caps
Travel End Caps should be cleaned after EVERY USE OR after
changing the o-rings by baking in an oven