background image

GATHER LINK SETUP INFORMATION

TEST PROCEDURES

12

 

Location   

Building or geographic area where the fiber link is located

 

Fiber Type 

Type of fiber used in the link under test; options may vary based on chosen cabling standard    

(see appendix for a fiber type diagram)

 

Injector End 

End of the fiber link where the light source unit will be used

RUN NAME – naming of fiber test results within the job to uniquely identify individual fiber strands

To make the setup process go more smoothly, have the following information ready in advance.  

LINK NAME – general information about the job (user-definable)

 

Meter End 

End of the fiber link where the meter unit will be used

 

Standard  

not required, since LOSS MODE does not determine PASS/FAIL

 TEST CORD TYPE – defines inter-connection loss at the very ends of the fiber link .where the test equipment connects in 

(patch panels, wall outlets) – Options: REFERENCE-grade or STANDARD-grade

Testing MPO cables uses a similar setup procedure to LOSS mode (see LOSS MODE - CONFIGURE JOB PARAMETERS) to set up 
the meter for MPO loss testing.

 

Project   

Name of the overall project

FIBER TEST MODE – how to go about testing the fiber link
 

Test Mode 

LOSS 

 

SOURCE PORT 

refers to the type of test cord attached to the light source

 

OPM PORT 

refers to the type of test cord attached to the optical power meter 

LINK INFORMATION – physical configuration of the link under test

 

Connections 

Number of inter-connections in the middle of the link under test (patch panels, other mating    

sleeves) NOT including the connections at the far end of the link

 

Splices   

Number of splices in the link under test; can be either fusion or mechanical splices

 

Reference Method 

1-jumper reference method

ENCIRCLED FLUX – is EF compliance required for this test: YES or NO (only required for multimode testing)
 

Consult cabling standard documentation or end user requirements to determine if EF compliance is required; if 
so, special mode controller cables will be required for setting the optical reference (aka “zeroing”).

 

Number   

 

The starting fiber strand number in the link.   The number will be automatically     

incremented as the test results are saved.

 

Name 

 

 

The name used to identify the group of individual fiber strands in the link

TESTING MPO CABLES

Summary of Contents for MPO

Page 1: ... INC COM Revision 1 0 Optical Wavelength Laboratories OWL N9623 Old Hwy 12 Internet OWL INC COM Whitewater WI 53190 Phone 262 473 0643 Optical Wavelength Laboratories OPERATIONS GUIDE MPO OPTICAL SWITCHES ...

Page 2: ...iring 16 Introduction Optical Ports 6 About This Manual 3 Precautions 5 Appendices Specifications 6 Before You Begin 3 Product Label 5 Description 4 OTDR Testing 7 VFL Testing 8 General Features 6 Light Sources 9 Power Meters 10 Test Procedures Applications Testing MPO Cables 11 Operation Maintenance Re charging the Device Battery 17 ...

Page 3: ...athered from reading printed literature carefully or by thoroughly reading suppliedoperationsmanuals 3 zeroing orreferencingofpowerlevels 4 determiningthecorrectlinkbudgettopassorfailby 2 properfiberoptictestprocedures FOTPs Completeunderstandingofeachoftheseissuesiscriticalforperformingproperfiberoptictests Throughout this manual you will find various symbols that assist with understanding the pr...

Page 4: ...own buttons allowing users to target specific MPO channels to be tested Channelnumberisindicatedbyabright easy to read7 segmentLEDdisplay TheMPOopticalportiscoveredbyashuttereddustcover inordertopreventdebrisfromcollectingintheMPOport The optical switches are powered by a re chargeable Lithium Polymer battery allowing for hours of continuous use on a complete 12 charge The battery is charged via s...

Page 5: ...se caution when working with any optical equipment High intensity fiber optic laser sources output potentially dangerous high energy invisible light and could cause serious irreparable damage to the eye Thus it is recommended to NEVER lookintotheconnectorportofalightsourceortheendofafiber On the back of each DataCenter 12 MPO Optical Switch are labels similar to the one shown below containing mode...

Page 6: ...luetooth enabledOWLtesters PowerButton powerON OFFthedevice GENERAL FEATURES DATACENTER 12 SERIES MPO OPTICAL SWITCH MODEL s MPO OSW M12 MPO OSW S12 6 Storage Temperature Dimensions Operating Temperature Weight Battery Type Battery Life Display Type GENERAL Lithium Polymer 7 segment LED 40 to 85 C 10 ounces 284 g 2 87 x 4 42 x 1 25 up to 12 hours continuous use 5 to 70 C Fiber Type Single fiber Po...

Page 7: ... want to include the far end MPO connector in the OTDR test attach another MPO adapter and MPO OTDR launchcable totheendoftheMPOcableundertest However basedonthepolarityoftheMPOcableundertest pleasebeawarethatthefibersmayormaynotbenumberedthesameon theoppositeendoftheMPOcable Takethisintoaccountwhentestingintheoppositedirection RepeatSteps4and5untilallfibersintheMPOcablehavebeentested IMPORTANT NO...

Page 8: ...Thisishelpfultovisuallyverifythepolarityofthecableundertest Seediagrambelowforcommonstandards basedMPO polarityarrangements For example when the MPO optical switch is set to Channel 1 the light will be seen on Fiber 1 for Type A Fiber 12 for Type B and Fiber2forTypeC STEP4 PresstheUParrowbutton tochangetothenextchannel MPO CABLE UNDER TEST 1 12 Key Up Key Down Type A Straight Key Up Key Up Type B ...

Page 9: ...efarend Seediagrambelowforcommonstandards basedMPOpolarityarrangements STEP4 PresstheUParrowbutton tochangetothenextchannel Testing12 fiberMPOcableswithaSingle portLightSourceandDataCenter12MPOOpticalSwitch STEP1 PowerONboththelightsourceandtheMPOSwitch MPO CABLE UNDER TEST 1 12 Key Up Key Down Type A Straight Key Up Key Up Type B Reversed Key Up Key Down Type C Pairs Flipped 1 2 3 4 5 6 7 8 9 10 ...

Page 10: ...rambelowforcommonstandards basedMPOpolarityarrangements For example when the MPO optical switch is set to Channel 1 the light will be seen on Fiber 1 for Type A Fiber 12 for Type B and Fiber2forTypeC STEP4 PresstheUParrowbutton tochangetothenextchannel MPO CABLE UNDER TEST 1 12 Key Up Key Down Type A Straight Key Up Key Up Type B Reversed Key Up Key Down Type C Pairs Flipped 1 2 3 4 5 6 7 8 9 10 1...

Page 11: ...Switch MPO OWL TEST SAVE LOSS FIBER 1 ACME CORP HQ WEST M COMPUTER ROOM I NETWORK HUB 3 Length 100m WAVE xxx 1310 EIA TIA 568C 3 CON 2 X INDOOR SM SPL 2 X ENTER 0 24 1310nm dB 0 23 1550nm dB Optical Switch MPO OWL Optical Source MPO OWL AUTO PINNED TO PINNED TYPE B PINNED TO PINNED TYPE B UNPINNED TO UNPINNED TYPE B MPO ADAPTER TYPE B NOTE for automatic channel switching to function properly the B...

Page 12: ...NFIGURE JOB PARAMETERS to set up themeterforMPOlosstesting Project Nameoftheoverallproject FIBERTESTMODE howtogoabouttestingthefiberlink TestMode LOSS SOURCEPORT referstothetypeoftestcordattachedtothelightsource OPMPORT referstothetypeoftestcordattachedtotheopticalpowermeter LINKINFORMATION physicalconfigurationofthelinkundertest Connections Number of inter connections in the middle of the link un...

Page 13: ...D TO PINNED TYPE B UNPINNED TO UNPINNED TYPE B MPO ADAPTER TYPE B SWITCH MPO SC SC MPO METER SOURCE Using your test reference cables please connect the meter and source as shown and press any key to proceed Optical Switch MPO OWL Optical Source MPO OWL AUTO PINNED TO PINNED TYPE B PINNED TO PINNED TYPE B UNPINNED TO UNPINNED TYPE B MPO ADAPTER TYPE B TEST LENGTH ZERO SAVE LOSS MPO OM3 MPO LOSS OWL...

Page 14: ... 4 0 00 0 00 0 5 0 00 0 00 0 6 0 00 0 00 0 7 0 00 0 00 0 8 0 00 0 00 0 9 0 00 0 00 0 10 0 00 0 00 11 0 00 0 00 12 0 00 0 00 N A BY MPO B EIA TIA 568C 3 CON 2 SPL 0 50 0um CS1 LEN 0m 1 HELP ENTER Optical Switch MPO OWL Optical Source MPO OWL AUTO PINNED TO PINNED TYPE B PINNED TO PINNED TYPE B MPO ADAPTER TYPE B TEST LENGTH ZERO SAVE LOSS Testing MPO LOSS OWL M ER R WA MPO dBm Power OPM OLS PORT 85...

Page 15: ...SPL 0 50 0um CS1 LEN 0m 1 HELP ENTER STORE TEST RESULTS TotestadditionalMPOcables repeatsteps2through5foreachMPOcabletotestanddocument To view the stored test results see WORKING WITH STORED DATA 5 PressSAVEtostorethetestresults thenmodifytherunnameifnecessary Optical Switch MPO OWL Optical Source MPO OWL AUTO PINNED TO PINNED TYPE B PINNED TO PINNED TYPE B MPO ADAPTER TYPE B TEST LENGTH ZERO SAVE...

Page 16: ...lly theOWLtestdevicewillbeabletoautomaticallycontroltheMPOchannelswitching TheindicatorLEDwillblinkslowlywhenswitchingtheMPOchannel STEP2 InitiatetheBluetoothpairingprocedureontheOWLdevice OPERATION MAINTENANCE ACTIVATING BLUETOOTH PAIRING 16 RequiredAccessories SCport 2 5mmcleaningswabsorone clickcleaner MPOport MPOone clickcleaner CompressedAir optional Fiberopticinspectionscope 400xmagnificatio...

Page 17: ...dustanddirtbuildupbyreplacingthedustcapsaftereachuse Warranty The MPO optical switch comes standard with a two year factory warranty which covers manufacturer defect and workmanshiponly MAINTENANCE AND CALIBRATION INFORMATION WARRANTY INFORMATION 17 RE CHARGING THE DEVICE BATTERY Toavoiddamagetotheunitorharmtotheuser onlyuseapprovedbatterychargers Batterychargerelectricalspecifications TheLithiumP...

Reviews: