USER MANUAL
System use
(Rev. 6)
Ow-RX
45
If digital radiography is selected, the exposure times are indicated in
Digital sensor
Size
Small
Normal
Large
Incisor
0.04
0.08
0.10
Canine
0.04
0.08
0.10
Premolar
0.05
0.09
0.11
Lower molar
0.07
0.11
0.14
Upper molar
0.08
0.14
0.19
Front bite-wing
0.04
0.08
0.10
Rear bite-wing
0.08
0.14
0.19
Table 2
The times for permanent phosphor sensors are provided in Table 3.
Permanent phosphors
Size
Small
Normal
Large
Incisor
0.07
0.10
0.14
Canine
0.07
0.10
0.14
Premolar
0.08
0.11
0.14
Lower molar
0.09
0.14
0.17
Upper molar
0.09
0.16
0.20
Front bite-wing
0.07
0.10
0.14
Rear bite-wing
0.09
0.16
0.20
Table 3
NOTE:
The times indicated in the tables are relative to the selection 65kV.
The times for the 60kV selection are obtained by multiplying the values
in the 65kV table by 1.45; they are multiplied by 0.7 for the 70kV
selection.
WARNING:
The times indicated in the tables are those set by the manufacturer.
It is appropriate that the user adapts the pre-set exposure times
according to local regulatory requirements (if any) and the type of
receptor used as described in paragraph 8.3.
A service technician is required to reset the default times.
Summary of Contents for Ow-RX
Page 1: ...Version July 2019 Rev 7 Ow RX User Manual 0051...
Page 2: ......
Page 4: ...USER MANUAL Revision history Ow RX Rev 6 THIS PAGE IS INTENTIONALLY LEFT BLANK...
Page 22: ...USER MANUAL Description Ow RX Rev 6 16 5 DESCRIPTION 5 1 Identification plates 5 1 2 6 4 3...
Page 73: ...USER MANUAL Maintenance Rev 6 Ow RX 67...
Page 74: ...Cod 6959902203_Rev 6 0051...