Performing a programmed exposure
If the previous examination was carried out in manual exposure mode,
press one of the size selection or anatomic selection keys to switch to
programmed exposure mode.
In programmed mode it is possible to change the size, type of tooth and
kV value.
Each time the "Size selection" key is pressed , indicated
acoustically, the Large patient / Normal patient / Small patient selection
changes.
To change the selection of the type of tooth, use the "Tooth anatomic
selection" key .
Each time this key is pressed the selection of the type of tooth changes in
rotation. This is signalled acoustically and shown on the display.
Based on the type of film that is selected, the pre-set times are provided
Film (F)
Size
Small
Normal
Large
Incisor
0.07
0.10
0.13
Canine
0.07
0.10
0.13
Premolar
0.09
0.13
0.17
Lower molar
0.10
0.15
0.20
Upper molar
0.13
0.19
0.26
Front bite-wing
0.07
0.10
0.13
Rear bite-wing
0.13
0.19
0.26
Table 1
NOTES:
These values are related to film type F.
The Ow-RX system can be programmed to use film with different
sensitivity levels; the programmed times vary depending on the film's
multiplicative factor. It is possible to request this setting from the
technician during installation.
Summary of Contents for Ow-RX
Page 1: ...Version July 2019 Rev 7 Ow RX User Manual 0051...
Page 2: ......
Page 4: ...USER MANUAL Revision history Ow RX Rev 6 THIS PAGE IS INTENTIONALLY LEFT BLANK...
Page 22: ...USER MANUAL Description Ow RX Rev 6 16 5 DESCRIPTION 5 1 Identification plates 5 1 2 6 4 3...
Page 73: ...USER MANUAL Maintenance Rev 6 Ow RX 67...
Page 74: ...Cod 6959902203_Rev 6 0051...