USER MANUAL
Technical data
(Rev. 6)
Ow-RX
27
Tube-head characteristics
Manufacturer
Owandy Radiology
Rated voltage
60-65-70 kV
p
(selectable)
Tubehead power
420 W
Total filtration
≥
2.5 mm Al eq. @ 70 kV
HVL (Half value Layers)
> 2 mm Al eq.
Transformer insulation
Oil bath
Interval between the exposures /
duty cycle
30 times the X-ray time/
1 : 30
Minimum focus to skin distance
20 cm (optional 30 cm cone)
Focus position
See Figure 6
Target angle
See Figure 7
X-ray diameter (@ 20cm focus)
≤
6 cm (35x45 mm + 25x35 mm +
20x30 mm optional)
Cooling
Convection
Leakage radiation at 1 metre
< 0.25 mGy/h
Technical factors for leakage radiation
70 kV, 6 mA, 1 s duty cycle
1 exposure every 30 seconds
Max specified energy input in 1 hour
720mAs @ 70kV, 2s
Figure 6: Tubehead focus position
Figure 7: Tubehead target angle
Focus position
.
20 cm
12.5°