background image

3

1

 

Mycare MP 50

Service

5.4

Shut down

ATTENTION

(see  chapter  “Safety  indications  and  dan-
gers due to non-compliance”)
Please  consult  the  chapter  “Units  used”
before attempting to open or take-apart the
filter.

5.5

Replacing the cartridges

Close  the  cold  water  supply  shut-off

valve or the water filter shut-off valve.

Open  the  faucet  to  remove  the  water

pressure.

Remove   the    Mycare   water     filter  :

Press  the  clamping  ring  of  the  tubing
adapter  back  and  remove  the  tubing
from  the  unit.  Remove  the  water  filter
from the wall-mounting.

Remove  the  flange  cover  upward  (see

fig. 11).

Use an allen key to loosen all four bolt

head screws.

Turn the filter housing to the left up to the

limit stop and pull off downward.

7 Only  every  2  years:  Remove  and

replace the profiled flange seal.
The  profiled  side  of  the  seal  must  be
pointing towards the flange (see fig. 10).
Failure to do so can lead to leakage and
water damage.

Remove the used replacement cartridge.
Clean  the  seat  of  the  cartridge  in  the

housing  using  the  disinfectant  cloth

provided. 
IMPORTANT: Do not touch any plastic

parts,  especially  the  inside  of  the  filter
housing, or the cartridge, with your bare
hands.

10 Insert the new replacement cartridge in

the built-in rotary flange and press in up
to the limit stop.

11 Insert the filter housing into the bayonet

opening of the built-in rotary flange, turn
to the right up to the limit stop.

12 Use an allen key to tighten all four bolt

head screws

13 Replace flange cover.
14 Insert   the   Mycare   water    filter    into

the  wall-mouting  brackets  and  connect
the tubing and adapters.

15 Open the cold water shut-off valve or the

water filter shut-off valve and follow the
the steps in the section “Commissioning”.

Always  keep  the  filter  elements  in  the

following order in flow direction:
1. activated carbon block filter cartridge

2. micro-filter cartridge!

Fig. 10: Built-in rotary flange

profiled flange seal

Select the tightening torque (approx
4 Nm)  so  that  the  seal  closes  and
the  water  filter  is  not  damaged  or
stressed/distorted!

Particles  of  activated  carbon  may

become dislodged from the cartrid-

ge during flushing. 

This  has  no  bearing  on  the  filter

function  and  only  underlines  the

efficiency  of  the  cartridge  type

used. 

Should these particles prove to be

an  optical  disturbance,  please

simply  rinse  the  activated  carbon

block  filter  cartridge  for  a  few

minutes  before  installing  the  mi-

cro-filtration  cartridge  at  the  next

required cartridge change.

Summary of Contents for Mycare MP50

Page 1: ...and safety information carefully before installing and putting the unit into service These must always be handed to the owner user Installation and operating instructions OWA Mycare MP50 Point of use...

Page 2: ...suitable for use in COLD water drinking applica tions up to a maximum water tempera ture of 30 C 86 F The point of use water filter operates on the basis of a 2 stage function principle 1st stage acti...

Page 3: ...transmit heat back through its respective plumbing Always ensure that this product is installed so that that the water tempe rature never exceeds a maximum of 30 C 86 F for all applications If in doub...

Page 4: ...formation on the Mycare so that it can be used in a safe economical and appropriate manner It contains fundamental information which must be observed during installation opera tion and maintenance Str...

Page 5: ...the general danger symbols given in this manual may result in the following risks among others failure of important functions of this filter danger to persons due to electrical and mechanical effects...

Page 6: ...id in the country of operation In addition to this all valid regulations governing health and safety in the country of use are to be adhered to at all times The Mycare water filter may be used only as...

Page 7: ...r the Mycare water filter is between 103 kPa 15 psi to 345 kPa 50 psi Under these pressure ranges the Mycare MP 50 water filter works most effectively 2 2 Information on hazard risks 2 2 1 Electrical...

Page 8: ...ty as close as possible to the point of use see fig 4 ATTENTION The water supply is to be turned off shut off before installation and may only be reopened once installation is fully completed Failure...

Page 9: ...he Mycare water filter can be installed in all standard commercial drinking water pipes Please shut down the incoming water supply before Installation is started Pay attention to the flow direction se...

Page 10: ...nce all the air has been vented see fig 7 7 Keep on flushing with at least 3 liters of water i e 10 minutes at rated flow or until the effluent is clear 8 Turn the faucet off 9 Set up the indicator fo...

Page 11: ...1 1 Mycare MP 50 Installation Fig 8 Installation cold water tubing tubing water filter wall mounting water filter shut off valve shut off valve brackets Shut off valves and tubing by others...

Page 12: ...icator for the next filter ch ange Move the indicator button on the flange cover to the next filter change require ment filter change every 2 6 months at the latest 7 Check for leaks both immediately...

Page 13: ...ing or the cartridge with your bare hands 10 Insert the new replacement cartridge in the built in rotary flange and press in up to the limit stop 11 Insert the filter housing into the bayonet opening...

Page 14: ...es flange cover built in rotary flange replacement cartridge micro filtration filter housing complete Allen key SW 5 used cartridge micro filtration used cartridge activated carbon bleed screw filter...

Page 15: ...rofiled flange seal is missing assembled wrongly or needs to be replaced every 2 years Check if the profiled flange seal is missing assembled wrongly or needs to be replaced The profiled side of the s...

Page 16: ...hol spiritus which may damage the unit 8 Warranty and services All products are supplied in accordance with the legal requirements of warranty legislation in their country of use Please note that fail...

Page 17: ...re min 103 kPa 15 psi Water pressure max 345 kPa 50 psi Cartridge replacement interval 3 months n i m l 2 e t a r w o l F 0 53 US gpm Total capacity per filter cartridge MF 2000 l1 528 US gal AC 2000...

Page 18: ...Mycare MP 50 8 1 Spare parts MP 50 10 Spare parts MP 50 10 1 2 4 3 B 5 6 7 8 C A A 9...

Page 19: ...2110040 B Spare parts set Filter hood consisting of pos 5 6 7 8 1 2110102 C Spare parts set Wall mounting bracket 1 2110065 2 MP 50 r e v o c e g n a l F 1 1 w e r c s d a e h t l o B 2 3 Replacement...

Page 20: ...ts are reserved for modifications as a result of technical progress or further developments Claims with regard to models or products are excluded 9990112 2015 02 OWA GmbH Postfach P O Box 100831 D 737...

Reviews: