If this error occurs during a monitoring, a software component running in the background
–
called the “Restart
tool”
will:
•
Close OsOne
•
Restart OsOne
•
Restart the monitoring
If the error is not automatically resolved after 3 consecutive tries of the “Restart tool”, monitoring is stopped.
How to fix it?
•
Check if any tubing is pinched
•
Check in the OsOne “Live” image if no ag
gregates are present in the field of view and if yes, try to flush
the fluidic system
•
Replace the BioConnect
How to prevent it?
•
Be careful not to pinch any tubing during the setup
•
Have the motor replaced during the maintenance if it appears that it is too old and makes noise during
normal operations.
Pump driver error
Description:
A hardware error related to the pump motor has been detected by OsOne
Cause(s):
There was an unexpected error with the pump:
When?
This alarm can be triggered anytime when OsOne uses the BioConnect
Type:
ERROR
Summary of Contents for ILine F Pro
Page 1: ...1 Ovizio Imaging Systems ILine F Pro User Manual...
Page 12: ...12 Introduction...
Page 16: ...16 Connect...
Page 17: ...17 Get Started Description BioConnect drawing...
Page 25: ...25 Press here to restart OsOne...
Page 26: ...26 Measure...
Page 39: ...39 Now you can disconnect the BioConnect pump from the Microscope pump motor...
Page 41: ...41 Troubleshooting...
Page 68: ...68 Configure...
Page 76: ...76...
Page 77: ...77...
Page 86: ...86 Remote Access...
Page 88: ...88 Maintenance and Servicing...