16. GENERAL SPECIFICATIONS
16.1 Sensor Specifications
Table 6.
ITEM
DESCRIPTION
Installation
Fixed type
Hottap type
Pipe Nominal Dia.
200, 250, 300, 400, 500, 600,
800, 000, 500, 2000mm
400, 500, 600, 800,
000, 500, 2000mm
Probe Nom. Bore
50mm
Materials
Meter Body
SUS304
Bluff Body
SUS304
Pressure Ratings
JIS 0K, ASME 50, JPI 50
Flange Rating
00mm (4") JIS 0K, ASME 50, JPI 50
Orientation
Horizontal or vertical pipeline
Horizontal pipeine
Max. Operating Pressure
Depending on flange rating
Operating Temp. Range
☆
−10 to 300°C
Fluid Velocities
Air (atm. pressure): 2 to 50 m/s
Water: 0.6 to 6 m/s
Standard Insertion Depth
Pipe nominal bore : D < 500mm ; 0.5D D = pipe nominal bore (mm)
Pipe nominal bore : D
≧
500mm ; 0.2D
Accuracy
★
Better than ±2% of full scale
Acceptable Fluid
Liquid (water, hot water, chemical liquids)
Gas (gases in general, compressed air, etc.)
Steam (saturated steam, superheated steam)
☆
: Operating temperature range varies according to fluid conditions.
★
: The stated accuracy is guaranteed only when used under the metering specifications (contributing factors
such as the pipe inside diameter and fluid conditions).