![Oval EX DELTA II Instructions Manual Download Page 26](http://html1.mh-extra.com/html/oval/ex-delta-ii/ex-delta-ii_instructions-manual_1662161026.webp)
D-583-2N-E
26
10. PARAMETER SETUP PROCEDURE
10.1 Procedure to modify a parameter
Given below are the steps to change current parameters to new ones: (This parameter modification procedure
uses MODE and RESET switches in the internal electronics. See Fig. . on page 3.)
() In "Measure Mode (normal mode)", hold MODE switch depressed for 5 seconds to go into "Review Mode".
⇩
(2) Using MODE and RESET switches, show the parameter you want to modify.
⇩
(3) Hold MODE switch depressed for 2 seconds to go into "Parameter Setup Mode".
⇩
(4) Using MODE and RESET switches, set up a new parameter. (See paragraph 2 below for detail procedure).
⇩
(5) Following the parameter entry, hold MODE switch depressed for 2 seconds to go back to "Review Mode".
⇩
(6) Using MODE and RESET switches, show the title (= one from bdAtA, AnA, PuLSE, or CorrEctCo).
⇩
(7) Hold MODE switch depressed for 5 seconds to go back to "Measure Mode".
A diagram to show parameter setup flow
Measure Mode
(normal mode)
Review Mode
Parameter
Setup
Mode
①
②
③
⑦
⑥
⑤
④
NOTE: For complete detail of MODE and RESET switch operations in (), (2), (6) and (7), see "Table 0.2
Menu Trees and Switch Operation" on page 29.
10.2 Procedure to Enter a Parameter
The procedure to enter a parameter (switch operations sequence in "Parameter Setup Mode") comes in three
ways (numerical setup, unit-of-measure setup, and decimal point location setup) as follows:
[Kind 1] Numerical setup parameters (F, H, Pu, At, A)
The blinking digit in the parameter setup mode is the place of interest.
MODE ....... Each time the switch is turned on, the place of interest moves one place to the left.
RESET ......
Each time the switch is turned on, the figure of interest increases by one, or the sign changes
(such as "E" or "-").
→
Following the parameter setup, hold the "MODE" switch depressed for 2 seconds (the new setting is then
established and the screen returns to the review mode).
Example: Parameter "F" (meter factor)
Digit of interest (blinking)
• Turning MODE "ON" moves the blinking digit one place to the left ("E").
• Turning RESET "ON" increases the figure by one ("2"
→
"3").
Exponential sign (E: 10
+n
, -: 10
-n
)
(The screen above reads F=1.2345 x 10
+2
L.)
L
(1)
(7)
(2)
(6)
(3)
(5)
(4)
Summary of Contents for EX DELTA II
Page 49: ...D 583 2N E 49...