![Otto Bock Terra Instructions For Use Manual Download Page 16](http://html.mh-extra.com/html/otto-bock/terra/terra_instructions-for-use-manual_1659304016.webp)
16
CAUTION
Altered centre of gravity resulting from use of the seat cushion
Risk of tipping over or falling due to unstable settings
►
After installing the seat cushion in the wheelchair, always check the user's sitting position
and the wheelchair's centre of gravity. The sitting position and centre of gravity must be
appropriate for the intended use and the user’s requirements. Work with the user to find the
optimal sitting position.
►
If necessary, adjust the position of the seat and axle to improve stability. Alternatively assess
the suitability of anti-tip devices.
CAUTION
Lack of testing prior to long-term use
Risk of skin reddening or pressure sore formation due to incorrectly selected seat cushion
►
Any seat cushion should be tried out by the user under medical supervision for several hours
prior to long-term use. The skin should be monitored for redness and/or other damage.
4.3 Safety instructions for use
WARNING
Improper handling of ignition sources
Risk of severe burns due to seat cushions, padding and upholstery catching on fire
►
The seat cushions, padding and covers fulfil the normative requirements for flame resist
ance. However, they may still ignite if fire is handled improperly or negligently.
►
Keep away from all ignition sources, especially lit cigarettes and open flames.
WARNING
Exceeding the service life
Serious injuries due to failure to observe the manufacturer’s requirements
►
Using the product beyond the specified expected service life leads to increased residual
risk.
►
Observe the specified service life.
WARNING
Independent modification of settings
Serious injuries to the user due to improper changes to the product
►
Do not modify the settings established by the qualified personnel. Only the settings
described in the section “Use” in these instructions for use may be adjusted independently.
►
In case of problems with the settings, please contact the qualified personnel who adjusted
your product.
CAUTION
Skin damage
Skin damage or pressure points due to overloading
►
Check your skin for intactness before and during use of the product.
►
Pay attention to diligent skin care and pressure redistribution during interruptions in using
the product.
►
If skin damage or other problems occur during use, stop using the product. Consult the
qualified personnel.