16
5) Depending on the size of the chosen Protector foot cuff, mark the following distances in at
least three places on the Protective Cover (Abmessen):
INFORMATION: The distances have to be measured from the foot contact surface
each time, creating a cut edge parallel to the foot contact surface. This allows the Pro
tector foot cuff to overlap over the Protective Cover main component to an ideal
extent.
• Size S: 110 mm
• Size M: 120 mm
• Size L: 125 mm
6) Remove the Protective Cover.
7) Fasten the upper closure (see fig. 14) and lengthwise closure (see fig. 7).
8) Tape off the Protective Cover with crepe adhesive tape above the mark.
9) Using a vibrating or band saw, shorten the Protective Cover along the bottom edge of the
crepe adhesive tape (see Fig. 8).
INFORMATION: Make sure that the closures are not shifted while sawing.
10) Deburr the cut edge.
NOTICE
Loss of stability due to shortening the Protective Cover too much
Protective Cover does not stay in place on the knee joint.
►
At the most, the Protective Cover may be shortened to just before the marked rib (see fig. 3).
Additional Instruction for Using a Torsion Adapter
►
Using suitable tools, remove the profile ribs on the interior of the Protective Cover (see
Fig. 9).
6.3 Installing the Protective Cover and Foot Cuff
6.3.1 Installing the Foot Cuff
1) Remove the prosthetic foot from the tube adapter.
In doing so, unscrew the two set screws that are screwed in the furthest but are not opposite
each other. Mark the two set screws.
2) Pull the foot cuff over the tube adapter (see Fig. 10).
3) Install the prosthetic foot on the tube adapter (see Fig. 11).
Apply Loctite® 241 to the two set screws and screw them in.
4) Position the clips of the foot cuff on the footshell, aligning the seam to posterior.
5) Attach the clips in the footshell (see Fig. 12).
INFORMATION: When using the 1C40 C-Walk, the clips that do not engage have to be
removed.
6.3.1.1 Installing the Protector foot cuff on the 1B1* Meridium prosthetic foot
Required Tools
Designation
Reference number
Fast acting adhesive
636K11, 636K36, Loctite® 406
When the product is used in combination with the 1B1* Meridium prosthetic foot, the following
steps have to be completed in addition:
Summary of Contents for C-Leg Protector 4X860 Series
Page 2: ...1 2 4 3 5 6 7 2...
Page 3: ...8 9 10 11 12 13 14 15 3...
Page 120: ...120 3 3 3 4 4 4 1 4 2 1 2 4 3...
Page 128: ...128 3 3 3 4 4 4 1 4 2 1 2 4 3...
Page 136: ...136 3 4 4 4 1 4 2 1 1 2 4 3 5 5 1 1 C Leg Protector 1 4P860 R 1 4P860 U...
Page 150: ...150 Shield Insert 3 3 3 4 4 4 1 4 2 1 2 4 3...
Page 155: ...155 Shield Insert 450 g 60 g Shield Insert 63 g 10 10 1 PPPP YYYY WW PPPP YYYY WW...