6A40 MagnoFlex Lock
Fig.
Item
Quantity Designation
Reference number
1 (
)
Set Screw (for ratchet unit)
506G3=M3X5-"Niro"
1
2
1 ( )
Lamination Adapter
6A43
1
3
1 ( )
Flexible Pin
6Y13=F1
1
4
1 (
)
Screw
503S3
1
5
1 ( )
Vacuum Forming Dummy
–
1
6
4 (
)
Snap Bushing
–
1
7
4 (
)
Cap Screw
501T28=M6X8
1
8
4 (
)
Flat Head Screw
501S41=M6X45
1
8
4 (
)
Flat Head Screw
501S41=M6X50
1
9
1 ( )
Ratchet Unit
5A52=K
5 Preparation for Use
CAUTION
Incorrect alignment or assembly
Injuries due to fall resulting from damaged prosthesis components
►
Observe the alignment and assembly instructions.
INFORMATION
Not all of the materials listed may be available in your country.
►
If the materials are not available, please contact the local branch of the
manufacturer to obtain information on alternative materials.
5.1 Fabricating the Prosthetic Socket
5.1.1 Preparing for Socket Fabrication
>
Required tools and materials:
6A43 Lamination Adapter, 501T28=M6X8 Cap Screws, Vacuum forming
dummy, 503S3 Screw, 633W8 Wax, 636K8=* Plastaband
1) Isolate the cap screws (thread and head from the outside) with wax.
2) Screw the cap screws into the lamination adapter (see fig. 2).
3) Fill the heads of the cap screws with Plastaband (see fig. 3).
4) Position the lamination adapter on the plaster positive along the longitud
inal axis of the residual limb.
5) Place the vacuum forming dummy into the round opening of the lamina
tion adapter and secure the two parts to the plaster positive with the flat
head screw (see fig. 4).
6) Isolate the head of the flat head screw with Plastaband.