background image

Ottobock  |  3

2R117

English

Last update: 2016-11-28
•  Please read this document carefully before using the product.
•  Follow the safety instructions to avoid injuries and damage to the product.
•  Instruct the user in the proper and safe use of the product.
•  Please keep this document in a safe place.

1 Indications for use  

The 2R117 Socket Connector connects the air path from the socket to the Harmony Pump. 

It can also be used to connect the 4R136 Ottobock One-Way Valves. 
The Harmony Socket Connector is to be used solely for the prosthetic fitting of the lower limb.

2 Components

The Harmony Socket Connector Kit includes the following components: 

Socket Connector

2R116

2

ea

Sandpaper

SL=P091-1

2

ea

Cotton tipped applicator 6”

640L8

2

ea

Black collar

SL=4Y310

2

ea

Instructions For Use

647G404

1

ea

3 Installation

CAUTION
Incorrect alignment or assembly
Injuries due to fall resulting from damaged prosthesis components
• Observe the alignment and assembly instructions.

INFORMATION
Failure to follow the installation instructions carefully could result in a complete loss of va-

cuum to the socket and thus a significant loss in suspension force.

To install the 2R117 Harmony Socket Connector, you will also need the following tools: 
• Needle nose pliers
• ¼" (6 mm) drill bit and drill
• Isopropyl alcohol
• Urethane adhesive (w/stick) (2 components) SL=P091
• Knife with curved blade
• Electrical tape
On a new socket, drill a ¼" (6 mm) hole into the socket at a distal, posterior position (Fig.1). 

For sockets with a 4R136 Ottobock One-Way Valve, remove valve from socket and drill into 

existing hole. Remove burrs from the inside of the socket.
Lightly sand both the interior and exterior of the socket with medium grit sandpaper to roughen 

the bonding surface. Wipe with isopropyl alcohol to clean the bonding surface. Also clean the 

2R117 Socket Connector with isopropyl alcohol. 

Summary of Contents for 2R117

Page 1: ...2R117 Instructions for use 3 Gebrauchsanweisung 5 ...

Page 2: ...2 Ottobock 2R117 6 4 5 6 mm 3 1 2 7 ...

Page 3: ...rect alignment or assembly Injuries due to fall resulting from damaged prosthesis components Observe the alignment and assembly instructions INFORMATION Failure to follow the installation instructions carefully could result in a complete loss of va cuum to the socket and thus a significant loss in suspension force To install the 2R117 Harmony Socket Connector you will also need the following tools...

Page 4: ...or Finished prosthesis shown in Fig 6 4 Legal information All legal conditions are subject to the respective national laws of the country of use and may vary accordingly 4 1 Liability The manufacturer will only assume liability if the product is used in accordance with the descriptions and instructions provided in this document The manufacturer will not assume liability for damage caused by disreg...

Page 5: ...Fehlerhafter Aufbau oder Montage Sturzverletzungen durch Schäden an Prothesenkomponenten Beachten Sie die Aufbau und Montagehinweise INFORMATION Werden die Montagehinweise nicht befolgt so kann dies einen kompletten Verlust des Un terdrucks zur Folge haben und damit auch einen signifikanten Unterschied in der Haftung Folgende Werkzeuge werden zusätzlich bei der Montage des Schaftansatzstückes benö...

Page 6: ...chieben Das Schlauchende auf das Lufteinlaßventil des Harmony stecken und den Sicherungsring entsprechend platzieren um die Verbindung zu sichern Das Klebeband entfernen Abb 6 4 Rechtliche Hinweise Alle rechtlichen Bedingungen unterliegen dem jeweiligen Landesrecht des Verwenderlandes und können dementsprechend variieren 4 1 Haftung Der Hersteller haftet wenn das Produkt gemäß den Beschreibungen u...

Page 7: ...Ottobock 7 2R117 ...

Page 8: ...Care LP 3820 West Great Lakes Drive Salt Lake City UT 84120 USA T 1 800 328 4058 F 1 800 962 2549 Otto Bock HealthCare GmbH Max Näder Straße 15 37115 Duderstadt Germany T 49 0 5527 848 0 F 49 0 5527 72330 healthcare ottobock de www ottobock com ...

Reviews: