The instrument and any device to be connected which has its own power supply should be turned off before any con-
nections are established.
•
We recommend that an annual calibration be performed on Alpha OAE and probe. Furthermore, we recommend that
calibration be performed if the equipment has suffered any potential damage.
•
For safety reasons, accessories connected to the equipment's outlet sockets must be supplied with the system.
•
Disposable accessories, such as eartips, should not be reused and must be replaced between patients to prevent cross-
infection.
•
This class of equipment is allowed in domestic establishments when used under the jurisdiction of a health care pro-
fessional.
•
Alpha OAE’s RF emissions are very low and are not likely to cause any interference in nearby electronic equipment,
but negative effect or loss of functionality of other local devices may occur if they are placed in close vicinity of Alpha
OAE.
•
When assembling an electro-medical system, the person carrying out the assembly must take into account that other
connected equipment which does not comply with the same safety requirements as this product may lead to a reduc-
tion in the overall safety level of the system.
•
In the United States of America, Federal law restricts this device to sale by or on the order of a licensed physician.
App. 1.5
Warranty
Alpha OAE is guaranteed against original defects in material and workmanship. It is also guaranteed to perform in accord-
ance with the manufacturer's specifications for a full 2 years from the date of purchase.
The warranty does not apply to the battery, to wear parts and disposable items.
This warranty only applies to those instruments purchased from authorized distributors or representatives. The purchaser
must return the instrument to an authorized distributor or representative and bear the cost of shipping.
This warranty does not cover breakage or failure due to tampering, misuse, neglect, accidents, modification or shipping
and is void if the instrument is not used in accordance with the manufacturer's instructions.
Please return your instrument to your local distributor or representative for warranty attention!
Important
•
The warranty will become void if the probe cleaning instructions are not strictly adhered to. If the
probe is not cleaned regularly as directed in the instructions, cerumen may result in blockage of the probe, where
sensitive components are contained and could be damaged.
App. 1.6
Repair, after-sales service and regular checks
If Alpha OAE as such is found to be defective or in some way varies from the manufacturer's specifications, an authorized
dealer will repair, replace or re-calibrate the instrument at no cost to the purchaser while Alpha OAE is within the war-
ranty period.
Service and repair of electro-medical equipment should only be carried out by the equipment manufacturer or by author-
ized representatives.
Otometrics - MADSEN Alpha OAE
53
App. 1 Standards and Safety
Summary of Contents for MADSEN Alpha OAE
Page 1: ...MADSEN Alpha OAE Screener User Manual Doc no 7 50 1370 US 01 Part no 7 50 13700 US 0 4 5 9 ...
Page 6: ...6 Otometrics MADSEN Alpha OAE ...
Page 10: ...1 Introduction 10 Otometrics MADSEN Alpha OAE ...
Page 16: ...2 Unpacking and Installation 16 Otometrics MADSEN Alpha OAE ...
Page 20: ...3 Getting Started with Alpha OAE 20 Otometrics MADSEN Alpha OAE ...
Page 32: ...5 Testing with Alpha OAE 32 Otometrics MADSEN Alpha OAE ...
Page 44: ...8 Handling Data in Alpha OAE 44 Otometrics MADSEN Alpha OAE ...
Page 50: ...9 Maintenance and Cleaning 50 Otometrics MADSEN Alpha OAE ...
Page 68: ......