9
Maintenance and Cleaning
9.1
The probe tip and probe body
Be sure to follow any established infection control procedures for the setting in which you are working.
Warning
•
Never place the probe tip in the ear canal without using a clean eartip.
The probe tip usually does not come into contact with the skin or secretion from the ear canal, as it is covered by the ear-
tip, but you should check the channels in the probe tip every time you have used the probe. Even small amounts of ceru-
men or vernix can block the probe channels or be deposited on the probe tip.
Warning
•
If needed, replace the probe tip with a spare one.
Note
•
Probe tips should be disposed of according to local regulations.
9.1.1
Cleaning and disinfecting the probe tip
Please remove the ear probe from the patient’s ear and separate the probe tip from the disposable ear tip before clean-
ing.
Caution
•
You should always comply with local hygienic standards.
See
.
The probe tip
–
Remove the probe tip from the probe.
Otometrics - MADSEN Alpha OAE
45
9 Maintenance and Cleaning
Summary of Contents for MADSEN Alpha OAE
Page 1: ...MADSEN Alpha OAE Screener User Manual Doc no 7 50 1370 US 01 Part no 7 50 13700 US 0 4 5 9 ...
Page 6: ...6 Otometrics MADSEN Alpha OAE ...
Page 10: ...1 Introduction 10 Otometrics MADSEN Alpha OAE ...
Page 16: ...2 Unpacking and Installation 16 Otometrics MADSEN Alpha OAE ...
Page 20: ...3 Getting Started with Alpha OAE 20 Otometrics MADSEN Alpha OAE ...
Page 32: ...5 Testing with Alpha OAE 32 Otometrics MADSEN Alpha OAE ...
Page 44: ...8 Handling Data in Alpha OAE 44 Otometrics MADSEN Alpha OAE ...
Page 50: ...9 Maintenance and Cleaning 50 Otometrics MADSEN Alpha OAE ...
Page 68: ......