RF emissions
CISPR11
Class A
ICS AirCal is suitable for use in hospital and clinic environments.
Note: The EMISSIONS characteristics of this equipment make it suitable for use in industrial areas
and hospitals (CISPR 11 class A). If it is used in a residential environment (for which CISPR 11 class
B is normally required) this equipment might not offer adequate protection to radio-frequency
communication services. The user might need to take mitigation measures, such as relocating or
re-orienting the equipment.
Harmonic emissions IEC
61000-3-2
Complies
Voltage fluctuations/flicker
emissions IEC 61000-3-3
Complies
Guidance and manufacturer's declaration - electromagnetic immunity for all equipment and systems
ICS AirCal is intended for use in the electromagnetic environment specified below. The user of ICS AirCal should ensure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance level
Electromagnetic environment - guidance
Electrostatic discharge (ESD)
IEC 61000-4-2
+/- 8 kV contact
+/- 2 kV, +/- 4 kV,
+/- 8 kV, +/- 15 kV air
+/- 8 kV contact
+/- 2 kV, +/- 4 kV,
+/- 8 kV, +/- 15 kV air
Floors should be wood, concrete or ceramic tile. If floors
are covered with synthetic material, the relative humid-
ity should be at least 30 %.
Electrical fast transient/burst
IEC 61000-4-4
+/- 2 kV for power supply lines
+/- 1 kV for input/output lines
+/- 2 kV for power supply lines
+/- 1 kV for input/output lines
Mains power quality should be that of a typical com-
mercial or hospital environment.
Surge
IEC 61000-4-5
+/- 1 kV line(s) to line(s)
+/- 2 kV line(s) to earth
+/- 2 kV I/O line(s) to earth
+/- 1 kV line(s) to line(s)
+/- 2 kV line(s) to earth
+/- 2 kV I/O line(s) to earth
Mains power quality should be that of a typical com-
mercial or hospital environment.
Voltage dips, short inter-
ruptions and voltage vari-
ations on power supply input
lines
IEC 61000-4-11
0 % U
T
; 0.5 cycle
At 0°, 45°, 90°, 135°, 180°, 225°, 270°
and 315°
0 % U
T
; 1 cycle
and
70 % U
T
; 25/30 cycles
Single phase: at 0°
0 % U
T
; 0.5 cycle
At 0°, 45°, 90°, 135°, 180°, 225°, 270°
and 315°
0 % U
T
; 1 cycle
and
70 % U
T
; 25/30 cycles
Single phase: at 0°
Mains power quality should be that of a typical com-
mercial or hospital environment. If the user of the ICS
AirCal requires continued operation during power
mains interruptions, it is recommended that the ICS
AirCal be powered from an uninterruptible power sup-
ply or a battery.
Voltage interruptions on
power supply input lines
IEC 61000-4-11
0 % U
T
; 250/300 cycles
0 % U
T
; 250/300 cycles
Power frequency
(50/60 Hz) magnetic field
IEC 61000-4-8
30 A/m
No relevant ports that could be
affected
Power frequency magnetic fields should be at levels char-
acteristic of a typical location in a typical commercial or
hospital environment.
U
T
is the AC mains voltage prior to application of the test level.
Guidance and manufacturer's declaration - electromagnetic immunity - for equipment and systems within Professional Healthcare use environment
ICS AirCal is intended for use in the electromagnetic environment specified below. The user of ICS AirCal should ensure that it is used in such an environment.
28
Otometrics - ICS AirCal
10 Notes on EMC (Electromagnetic Compatibility)