12
AS7050 PC Software
Starting a ECG Raw Data measurement
Fig. 10: Overview of the GUI (EVK measurement)
1.
Firstly, press the sensor button
”S
1
”
for 1 second
to power on the sensor.
2.
Afterward, select the correct
COM port number
or
BLE MAC address
from the drop-down menu.
3.
Then, click the
Connect
button.
4.
When it is successfully connected, the button on
the left-hand side of the status bar (AS7050) will
change color to green
(COM connection)
and blue
(BLE connection)
.
5.
The green LED on the Bluetooth module will start
blinking as soon as the board connects with the
GUI. In the case of a BLE connection, the LED will
blink blue instead.
6.
After connecting, select the correct configuration
file from the configuration presets provided (
EVK:
ECG 200 Hz, Gain 256
using Electrodes) in the
configuration tab.
7.
Then, click the
Start Measurement
button.
8.
Finally, place three fingers on the ECG electrodes,
and subsequently, you will see the ECG raw data
plot in the Graph window.