oserio FTG-315 series Instruction Manual Download Page 7

10

11

Before Using the Monitor

1.7 EMC Guidance and Manufacturer's Declaration

The FTG-315 is intended for use in the electromagnetic environment specified below. The customer or the user of the 
FTG-315 should assure that it is used in such an environment. 

Guidance and manufacturer’s declaration-electromagnetic immunity

Guidance and manufacturer’s declaration-electromagnetic immunity

Immunity test

IEC 60601 test level

Compliance level

Electromagnetic environment-guidance

Immunity test

IEC 60601 test level

Compliance level

Electromagnetic environment-guidance

Electrostatic 
discharge(ESD)
IEC 61000-4-2

Electrical fast 
transient/burst 
IEC 61000-4-4

± 6 kV contact
± 8 kV air

± 6 kV contact
± 8 kV air

Floors should be wood, concrete or ceramic 
tile. If floors are covered with synthetic 
material, the relative humidity should be at 
least 30%.

± 2kV for power supply 
lines + 1kV for 
input/output lines

± 2kV for power 
supply lines 
Not applicable

Mains power quality should be that of a 
typical commercial or hospital environment.

Surge 
IEC 61000-4-5

± 1kV line(s) to line(s) 
± 2kV line(s) to earth

± 1kV differential 
mode 
Not applicable

Mains power quality should be that of a 
typical commercial or hospital environment.

Voltage Dips, 
short interrupt-
ions and voltage
variations on 
power supply 
input lines 
IEC 61000-4-11

<5% UT(>95% dip in UT)
for 0.5 cycle 
40% UT(60% dip in UT)
for 5 cycles 
70% UT(30% dip in UT)
for 25 cycles 
<5% UT(>95% dip in UT)
for 5 s

Mains power quality should be that of a 
typical commercial or hospital environment. 
If the user of the

 FTG-315, 

requires 

continued operation during power mains 
interruptions, it is recommended that the 

FTG-315

 be powered from an uninterruptible 

power supply or a battery.

Not applicable 

Not applicable 

Not applicable 

Not applicable

Power frequency
(50/60 Hz) 
magnetic field 
IEC 61000-4-8

3 A/m

3 A/m

The

 FTG-315

 power frequency magnetic 

fields should be at levels characteristic of a 
typical location in a typical commercial or 
hospital environment.

NOTE: UT is the a.c. mains voltage prior to application of the test level.

Summary of Contents for FTG-315 series

Page 1: ...of Health Management Products This Instruction Manual will guide you through the setup procedures and outline the key features Please keep the instruction manual at hand all the time for future refer...

Page 2: ...sed Symbol Description 1 7 EMC Guidance and Manufacturer s Declaration 1 2 3 4 5 8 9 Operating Instructions 2 1 Measuring Body Composition 2 2 Error Message List More Information 3 1 Care and Maintena...

Page 3: ...otal cholesterol and serum lipoprotein in children and adolescents American journal of public health After compared measurement value indicator will shows Low Normal High Very High 3 Large LCD display...

Page 4: ...asurement values Recommended Measurement Times It is recommended to use this unit in the same environment and daily circumstances 1 People with an electronic medical implant such as a pacemaker should...

Page 5: ...lean the electrodes and glass cleaner applied to a cloth rst to keep them shiny 16 Place the unit on a hard at surface where there is minimal vibration to ensure safe and accurate measurement 1 5 Impo...

Page 6: ...buildings used for domestic purposes Harmonic emissions IEC 61000 3 2 Voltage uctuations icker emissions IEC 61000 3 3 Not applicable Not applicable Before Using the Monitor 1 6 Used Symbol Descripti...

Page 7: ...ot applicable Mains power quality should be that of a typical commercial or hospital environment Surge IEC 61000 4 5 1kV line s to line s 2kV line s to earth 1kV di erential mode Not applicable Mains...

Page 8: ...AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess the electromagnetic environment due to xed RF transmitters an electromagnetic site survey should be...

Page 9: ...ding to the maximum output power of the communications equipment Rated maximum output power of transmitter W For transmitters rated at a maximum output power not listed above the recommended separatio...

Page 10: ...play user s weight 2 Run o care APP and enter bubble 4 Measuring Body Composition 2 1 Measuring Body Composition Note If you didn t step on the scale but LED bar is ashing please await till the LED tu...

Page 11: ...tenance 1 Store the unit in a safe and dry location 2 Avoid place on the aisle to avoid accidentally tripped over pedestrians 3 Never submerge in water 4 Do not expose the unit to direct sunlight extr...

Page 12: ...ironment Limit Temperature 5 C to 35 C Humidity 30 to 85 RH Temperature 20 C to 60 C Humidity 10 to 95 RH Maximum atmospheric pressure range Maximum operation altitude 1013hpa 800hpa 2000m Scale Displ...

Reviews: