OrthAlign KneeAlign 2 System Technique Manual Download Page 4

KneeAlign

®

 2 System

 Surgical Technique Manual

KneeAlign

®

 2 System

 Surgical Technique Manual

4

Set up

Set up

Assemble tibial Jig

 •   

Assemble tibial jig as follows:

 •   

Set fixation arm on proximal end of tibial jig to “Left”  

or “Right” for leg being operated on.

 •   

Ensure that varus/valgus and posterior slope levers  

of tibial jig are locked.

 •   

Attach midline probe 

(a)

 to proximal end of tibial jig.

 •   

Attach shin spacer 

(b)

 to tibial jig.

 •   

Attach ankle tube 

(c)

 to distal end of tibial jig and lock  

in place.

 •   

Attach malleolar probe 

(d)

 to ankle tube.

 •   

Press center button on navigation unit when completed.

TIP:   

Fixation arm is spring-loaded and may be adjusted by 

pulling arm away from jig, then rotating.

TIP:  

Shin spacer may already be attached.

TIP:   

Squeeze button on ankle tube and insert malleolar probe 

through slot until it reaches limit of its travel.
Use curved malleolar probe to extend range for extremely  

long or short tibias.

Assemble Femoral Jig 

(cont.)

 •   

Once assembled, ensure microblock is in neutral position  

by adjusting varus/valgus and flexion/extension screws  

with ball driver.

 •   

Press center button when completed.

TIP:   

Plate and bracket of microblock should be approximately 

parallel to each other when viewed from front and side.

Step

Step

7

8

5

Assemble Femoral Jig

a  

Push guide rod in direction indicated by arrow labeled 

“Lock” until it stops.

b  

Fully extend slider to its maximum setting in order to 

position guide rod against gold latch. Ensure gold latch  

is unlocked.

c  

Lock guide rod by sliding gold latch into position.

d  

Slide femoral cutting block onto guide rod in direction 

indicated by arrow on cutting block labeled “Assemble”  

until it locks into place.

Lock

Maximum 

Setting

Assembled Tibial Jig

Tibial Jig Assembly Order

b

a

d

c

Neutral Position

F/e

V/V

Plate

Bracket

a

d

c

b

Summary of Contents for KneeAlign 2 System

Page 1: ...ite 500 Aliso Viejo CA 92656 866 582 0879 www orthalign com OrthAlign is committed to providing surgeons with user friendly cost effective surgical navigation products for precise alignment For more information about the KneeAlign 2 System please contact us at 866 582 0879 or info orthalign com About OrthAlign Inc ...

Page 2: ...rm Sensor ID 1 Step 3 Attach Bracket 2 Step 4 Hold Horizontally 2 Step 5 HoldVertically 3 Step 6 Verify Calibration 3 Step 7 Assemble Femoral Jig 4 Step 8 AssembleTibial Jig 5 Step 9 Select Knee 6 Step 10 Select Bone 6 FEMUR Step 1 Secure on Femur 7 Step 2 Input A P Offset 7 Step 3 Attach Sensors 8 Step 4 Maneuver Leg 8 Step 5 Set Resection Plane 9 Step 6 Set Resection Depth 10 Step 7 Finished Fem...

Page 3: ...ep 1 Rest navigation unit horizontally on level surface with screen facing upwards and hold steady with ball in circle Navigation unit will beep twice and automatically proceed to next screen TIP During calibration steps Navigation unit must be maintained in steady position or calibration will not be accepted A red hand will flash on screen to indicate that unit is not steady Level must be green b...

Page 4: ... slot until it reaches limit of its travel Use curved malleolar probe to extend range for extremely long or short tibias Assemble Femoral Jig cont Once assembled ensure microblock is in neutral position by adjusting varus valgus and flexion extension screws with ball driver Press center button when completed TIP Plate and bracket of microblock should be approximately parallel to each other when vi...

Page 5: ...teel headed pins provided in instrument tray should be used Do not impact or hammer system components TIP Verify that there is approximately 10 mm of clearance between femoral cutting block and anterior condyles Input A P Offset Read anterior posterior offset position of center pin on microblock slider scale Enter offset value into navigation unit by pressing right button until correct number is d...

Page 6: ...esults it is important that hip is kept stationary during registration Knee must be returned to start position within 2 cm and 5 rotation TIP Following maneuver completion during hip point calculation navigation unit screen will provide user with feedback on speed of maneuver in medial lateral M L and anterior posterior A P planes Green bars displayed for a given plane indicate that speed of the m...

Page 7: ...istal guide Ensure that distal guide paddles remain flush on highest point of condyle being referenced TIP When pinning the cutting block take care to avoid applying excessive force that could alter intended varus valgus and flexion extension angles Topreventpotentialinterference ofoblique pin with microblock pins donotinsertobliquepinuntilmicroblockhasbeenremoved Useangelwingstylustoverifyresecti...

Page 8: ...for pinning KneeAlign 2 instrumentation TIP If jig appears in extension relative to tibia remove shin spacer This may be the case for very obese legs Secure on Tibia a Pull malleolar probe to maximum anterior position in order to avoid interference during placement of jig b Place tibial jig on anterior tibia aligning etched line on fixation arm of jig with medial third of tibial tubercle c Positio...

Page 9: ...ith side levers TIP Malleolar probe may be adjusted to assist in setting posterior slope angle Do not use malleolar probe to set posterior slope angle if posterior slope lever is locked After locking levers do not move jig relative to the tibia Pushing or pulling locked bars of the jig may introduce errors in navigation accuracy KneeAlign 2 System does not display measurement values when oriented ...

Page 10: ...irm that posterior slope and varus valgus angles still match desired values Use angel wing stylus to confirm resection level and perform a visual confirmation of alignment TIP While setting resection depth stylus end should not be pressed into tibial plateau TIP Resection depth measurement should be set according to implant manufacturer s recommendations TIP If additional support is required place...

Page 11: ...ess center button to repeat maneuver at higher speed sufficient to obtain green color bars for both A P and M L speeds Sensor signal is lost Remove and reattach sensor battery to re establish connection Hip point registration maneuver not finished prior to end of navigation unit beeping stream Press center button to repeat maneuver finishing maneuver and returning knee to starting position prior t...

Page 12: ...ock Assembly 402045 Malleolar Probe Curved 401573 Malleolar Probe Straight 401572 Microblock 402042 Midline Probe Assembly 401571 Pin Driver Tri Flat 402067 Reference Sensor 2 133632 Resection Stylus Adjustable Captured 401587 Shin Spacer 401548 Threaded Pin Headed 46 mm 2 Pack 402395 Threaded Pin Headless 65 mm 2 Pack 402394 Tibial Cutting Block Left 401579 Tibial Cutting Block Right 401578 Curve...

Reviews: