OrthAlign KneeAlign 2 System Technique Manual Download Page 12

KneeAlign

®

 2 System

 Surgical Technique Manual

KneeAlign

®

 2 System

 Surgical Technique Manual

20

note:

 For cleaning and sterilization instructions, please refer to the Instructions for Use.

InStrument SpeCIFICAtIOnS

OrDerIng InFOrmAtIOn

Product description  

 

catalog number

KneeAlign® 2 Navigation Unit  

133631

KneeAlign® 2 instrument Set Replacement Parts

KneeAlign® 2 Tibial Jig Body 

401594

KneeAlign® 2 Mounting Bracket 

402046

Ankle Tube 

401582

Ball Driver (2.5 mm) 

402048

Calf Strap 

401592 

Distal Guide Assembly 

402043

Femoral Cutting Block Assembly 

402045

Malleolar Probe, Curved 

401573

Malleolar Probe, Straight 

401572

Microblock 402042
Midline Probe Assembly 

401571

Pin Driver, Tri-Flat 

402067

Reference Sensor 2 

133632

Resection Stylus, Adjustable, Captured 

401587

Shin Spacer 

401548

Threaded Pin, Headed (46 mm, 2-Pack) 

402395

Threaded Pin, Headless (65 mm 2-Pack) 

402394

Tibial Cutting Block, Left 

401579

Tibial Cutting Block, Right  

401578

Cur

ved M

alleolar 

Pr

obe

Tibial C

utting 

Blocks 

 

(lef

t and r

ight)

Sensor

Ank

le 

 

Tube

Calf 

 

Strap

Straight  M

alleolar  Pr

obe

Shin 

 

Spacer

Thr

eaded P

in

 

Headed 

 

(46 mm)

KneeAlig

n® 2

 

Tibial Jig Body

Pin 

 

Dr

iv

er

A

djustable  St

ylus

Ball 

 

Dr

iv

er

Distal 

 

Guide

M

icr

oblock

M

idline  Probe

M

ounting 

 

Brack

et

Distal  Femoral  Cutting  Block

Thr

eaded P

in

 

Headless 

 

(65 mm)

Summary of Contents for KneeAlign 2 System

Page 1: ...ite 500 Aliso Viejo CA 92656 866 582 0879 www orthalign com OrthAlign is committed to providing surgeons with user friendly cost effective surgical navigation products for precise alignment For more information about the KneeAlign 2 System please contact us at 866 582 0879 or info orthalign com About OrthAlign Inc ...

Page 2: ...rm Sensor ID 1 Step 3 Attach Bracket 2 Step 4 Hold Horizontally 2 Step 5 HoldVertically 3 Step 6 Verify Calibration 3 Step 7 Assemble Femoral Jig 4 Step 8 AssembleTibial Jig 5 Step 9 Select Knee 6 Step 10 Select Bone 6 FEMUR Step 1 Secure on Femur 7 Step 2 Input A P Offset 7 Step 3 Attach Sensors 8 Step 4 Maneuver Leg 8 Step 5 Set Resection Plane 9 Step 6 Set Resection Depth 10 Step 7 Finished Fem...

Page 3: ...ep 1 Rest navigation unit horizontally on level surface with screen facing upwards and hold steady with ball in circle Navigation unit will beep twice and automatically proceed to next screen TIP During calibration steps Navigation unit must be maintained in steady position or calibration will not be accepted A red hand will flash on screen to indicate that unit is not steady Level must be green b...

Page 4: ... slot until it reaches limit of its travel Use curved malleolar probe to extend range for extremely long or short tibias Assemble Femoral Jig cont Once assembled ensure microblock is in neutral position by adjusting varus valgus and flexion extension screws with ball driver Press center button when completed TIP Plate and bracket of microblock should be approximately parallel to each other when vi...

Page 5: ...teel headed pins provided in instrument tray should be used Do not impact or hammer system components TIP Verify that there is approximately 10 mm of clearance between femoral cutting block and anterior condyles Input A P Offset Read anterior posterior offset position of center pin on microblock slider scale Enter offset value into navigation unit by pressing right button until correct number is d...

Page 6: ...esults it is important that hip is kept stationary during registration Knee must be returned to start position within 2 cm and 5 rotation TIP Following maneuver completion during hip point calculation navigation unit screen will provide user with feedback on speed of maneuver in medial lateral M L and anterior posterior A P planes Green bars displayed for a given plane indicate that speed of the m...

Page 7: ...istal guide Ensure that distal guide paddles remain flush on highest point of condyle being referenced TIP When pinning the cutting block take care to avoid applying excessive force that could alter intended varus valgus and flexion extension angles Topreventpotentialinterference ofoblique pin with microblock pins donotinsertobliquepinuntilmicroblockhasbeenremoved Useangelwingstylustoverifyresecti...

Page 8: ...for pinning KneeAlign 2 instrumentation TIP If jig appears in extension relative to tibia remove shin spacer This may be the case for very obese legs Secure on Tibia a Pull malleolar probe to maximum anterior position in order to avoid interference during placement of jig b Place tibial jig on anterior tibia aligning etched line on fixation arm of jig with medial third of tibial tubercle c Positio...

Page 9: ...ith side levers TIP Malleolar probe may be adjusted to assist in setting posterior slope angle Do not use malleolar probe to set posterior slope angle if posterior slope lever is locked After locking levers do not move jig relative to the tibia Pushing or pulling locked bars of the jig may introduce errors in navigation accuracy KneeAlign 2 System does not display measurement values when oriented ...

Page 10: ...irm that posterior slope and varus valgus angles still match desired values Use angel wing stylus to confirm resection level and perform a visual confirmation of alignment TIP While setting resection depth stylus end should not be pressed into tibial plateau TIP Resection depth measurement should be set according to implant manufacturer s recommendations TIP If additional support is required place...

Page 11: ...ess center button to repeat maneuver at higher speed sufficient to obtain green color bars for both A P and M L speeds Sensor signal is lost Remove and reattach sensor battery to re establish connection Hip point registration maneuver not finished prior to end of navigation unit beeping stream Press center button to repeat maneuver finishing maneuver and returning knee to starting position prior t...

Page 12: ...ock Assembly 402045 Malleolar Probe Curved 401573 Malleolar Probe Straight 401572 Microblock 402042 Midline Probe Assembly 401571 Pin Driver Tri Flat 402067 Reference Sensor 2 133632 Resection Stylus Adjustable Captured 401587 Shin Spacer 401548 Threaded Pin Headed 46 mm 2 Pack 402395 Threaded Pin Headless 65 mm 2 Pack 402394 Tibial Cutting Block Left 401579 Tibial Cutting Block Right 401578 Curve...

Reviews: