background image

– 3 –

ORO-PULSE OXIMETER USER’S MANUAL

SECTION 1: SAFETY

1.1  Instructions for the Safe Operation and Use of the Pulse Oximeter

•  Do  not  attempt  to  service  the  pulse  oximeter.  Only  qualified 

service personnel should attempt any needed internal servicing.

•  Do not use the oximeter in situations where alarms are required. 

•  Prolonged use or the patient’s condition may require changing the 

sensor site periodically. Change sensor site and check skin integri

-

ty, circulatory status and correct alignment at least every 2 hours.

•  SpO

2

 measurements may be adversely affected in the presence of 

high ambient light. Shield the sensor area (with a surgical towel, or 

direct sunlight, for example) if necessary.

•  The following reasons will cause interference.

•  High-frequency electrosurgical. 

•  Placement of a sensor on an extremity with a blood pressure cuff 

arterial catheter, or intravascular line. 

•  The patient has hypotension severe vasoconstriction severe ane

-

mia or hypothermia. 

•  The patient is in cardiac arrest or is in shock. 

•  Fingernail polish or false fingernails may cause inaccurate SpO

2

 

readings. 

1.2  Warnings 

WARNING:

  EXPLOSION  HAZARD  —  Do  not  use  the  oximeter 

in  a  flammable  atmosphere  where  concentrations  of  flammable 

anesthetics or other materials may occur.

WARNING:

 Do not throw batteries in fire as this may causes them to 

explode.

WARNING:

 Do not use the pulse oximeter in an MRI or CT environ

-

ment.

CAUTION:

 Keep the operating environment free of dust, vibrations, cor

-

rosive, or flammable materials, and extremes of temperature and humidity.

CAUTION:

 Do not operate the unit if it is damp or wet because of 

condensation or spills. Avoid using the equipment immediately after 

moving it from a cold environment to a warm, humid location.

WARNING:

 Do not attempt to recharge normal dry-cell batteries, they 

may leak. And may cause a fire or even explode.

CAUTION:

 Never use sharp or pointed objects to operate the front

-

-panel switches.

CAUTION:

 The battery must be taken out from the battery compart

-

ment if the device will not be used for a long time.

CAUTION:

 The device shall only be used if the battery cover is closed.

CAUTION:

  The  battery  must  be  proper  disposed  according  to  local 

regulation after their use.

1.3  Definitions and Symbols

Symbol

Description

Symbol

Description

Type BF Equipment

Information of manufacture, 

including name and address

Refer to the instruction 

manual /booklet

When the end-user wishes 

to discard this product, it 

must be sent to separate 

collection facilities for 

recovery and recycling

Serial NO*

Note

The important information 

you should know

Caution

The information you should 

know to protect the equip

-

ment from possible damage

Warning

The information you should 

know to protect patients 

and medical staff from 

possible injury

SECTION 2: INTRODUCTION

2.1  Brief Device Description

The Pulse Oximeter, based on all digital technology, is intended for noni

-

nvasive spot-check measurement of functional oxygen saturation of arte

-

rial hemoglobin (SpO

2

). Advanced DSP algorithm can minish the influence 

of motion artifact and improve measurement accuracy of low perfusion.

The Oximeter can be used to measure human Hemoglobin Saturation 

and heart rate through finger. 

The product is suitable for family, Oxygen Bar, social medical organi

-

zations, physical care in sports and etc. in the case of saturation below 

80%, contact your doctor.

2.2  Intended Use

The product is dedicated for unprofessional user and can be used in 

house conditions.

2.3  Product Features

•  Lightweight for carrying and Easy-To-Use. 

•  Manually adjust the direction of interface .

•  OLED display, simultaneous display for testing value and plethy

-

smogram.

•  Visual alarm function. Real-time spot-checks.

•  Low Battery voltage indicator.

•  Automatically standby or sleep.

SECTION 3: INSTALLATION, SETUP, AND OPERATION

3.1 Description of the Front Panel

Item

Name

Description

1

OLED Panel

Display the SpO

2

/PR data & Bargraph, Plethysmogram,PI

2

Key

Start the working state

3

Battery 

 

Compartment

2*AAA 1.5V Alkaline battery

3.2   Display

After switch on, the OLED display is as follows:

3.3   Parameter setting

Press the direction button (>0.5s), the oximeter will enter into parame

-

ter seting. There are two submenu for choice:

•  When the „*” signal is show on the „Sounds setup”, press the but

-

ton (>0.5s), and enter into the sounds setting menu(Figure a), you 

can  press  the  button  in  turn  to  select  the  item,  and  press  the 

button to change the data you need. Select „+” or „-” to increase 

or decrease the number of settings.

•  When the „*” signal is show on the „Alm setup”, press the button 

(>0.5s), and enter into the alarm setting menu(Figure b). Press the 

button to set on/off the alarm and beep.

When the „* ” signal is show on the „Restore”, press the button (>0.5s) 

and or the settings are back to the factory settings.

Settings

Alm Setup  *

Alm 

off

Beep 

off

Demo 

off

Restore 

OK

Brightness  4

Exit

Settings

Sounds Setup  *

SpO

2

 Alm Hi  100

SpO2 Alm Lo  94

PR Alm Hi 

130

PR Alm Lo 

50

+/- 

+

Exit

Figure A

Figure B

3.4   Operation

 

Install battery

Installing two AAA batteries into battery cassette in correct polarities 

and cover it (as Figure)

Battery indicator

SpO

2

PR

Key
Direction

Indeks perfuzji

Pletyzmogram

Wykres słupkowy

Reviews: