EPSILON CLOCK MODEL EC20S
Spectracom
User’s Manual
48
Monitored status of internal 1PPS generated from oscillator frequency:
- OK
1PPS driver operational.
- Alarm
1PPS driver failure.
13.3.2.2
SYNCHRONIZATION
12.
PHASE ACCURACY
Status of the estimated phase accuracy during a holdover period:
- OK
The phase accuracy is below the Alarm Phase Threshold (see
13.7.3).
- Alarm
The phase accuracy is above the Alarm Phase Threshold (see
13.7.3).
13.
FREQUENCY ACCURACY
Status of the estimated frequency accuracy during a holdover period:
- OK
The frequency accuracy is below the Alarm Frequency Threshold
(see 13.7.3).
- Alarm
The frequency accuracy is above the Alarm Frequency Threshold
(see 13.7.3).
13.3.2.3
GNSS SOURCE
14.
GNSS LOCK
Performance of the 1PPS signal of the GNSS receiver:
- OK
1PPS signal of the GNSS receiver can be used as 1PPS reference
signal for disciplining operation.
- Alarm
1PPS signal of the GNSS
receiver can’t be used as 1PPS
reference signal for disciplining operation.
15.
ANTENNA STATUS
State of the connection to the GNSS antenna: