Service of the  
polarographic  
oxygen sensor 
OroboPOS

 

 

Gnaiger E 

 

OROBOROS INSTRUMENTS Corp

 

high-resolution respirometry 

Schöpfstr 18, A-6020 Innsbruck, Austria 

Email: 

[email protected]

 

www.oroboros.at 

 

 

 

 

Section 

1.  Accessories ............................................. 20  

Page

 

 

2.  Cleaning ................................................. 20 

3.  Membrane mounting ................................ 22 
4.  Electrical cable connection ........................ 23 

5.  Storage of the POS .................................. 23 
6.  Replacement of the POS head ................... 24 

7.  References .............................................. 24

 

 

 

 

Summary:

 

Service  of  the  polarographic  oxygen  sensors  (

OroboPOS,

 

POS

) is the basis for  signal stability, low noise and high time resolution. 

Sensor  service  may  not  be  required  for  several  months  of  operation.  
Performance  of  the  OROBOROS  Oxygraph-2k  according  to  instrumental 
specifications is obtained only with oxygen sensors  which are maintained 

in a good functional state. 

 

POS service is required if (1) a new 

sensor  is  prepared,  (2)  the  raw 
signal at air calibration declines, (3) 
signal  noise  is  high,  (4)  the  time 
response 

is 

prolonged 

(time 

constant  >10  s)  and  biphasic,  (5) 

the oxygen signal at zero calibration 
does  not  decline  rapidly  to  a  low 

level of zero current (0% to max. 5%). For each sensor, the frequency of 
POS  service  can  be  optimized  on  the  basis  of  a  long-term  record  of 

calibration values as a quality control (O2k-Calibration-List.xls). 

Mitochondrial Physiology Network 19.01(B01): 19-24 (2014) 

2014 OROBOROS

 

B:

  

www.bioblast.at/index.php/MiPNet19.01B_POS-Service

 

 

Version B01: 2014-03-06 

Oxygraph-2k Manual 

Summary of Contents for Oxygraph-2k

Page 1: ...s of operation Performance of the OROBOROS Oxygraph 2k according to instrumental specifications is obtained only with oxygen sensors which are maintained in a good functional state POS service is required if 1 a new sensor is prepared 2 the raw signal at air calibration declines 3 signal noise is high 4 the time response is prolonged time constant 10 s and biphasic 5 the oxygen signal at zero cali...

Page 2: ...Net14 01 For sensor service remove the black POS seal tip 1 It is normal to see many small bubbles in the electrolyte reservoir This does not indicate that this caused a problem while the sensor was actually in use Remove the PEEK membrane ring 2 3 and membrane 4 Fig 2 Wash off the electrolyte with distilled water For cleaning of the anode and cathode the sensor head is best removed from the Orobo...

Page 3: ...ig 3 Hold the sensor in a vertical position and polish the cathode for one minute in a figure eight motion Fig 3 Wash the aluminum oxide powder carefully off the sensor with distilled water and wash the cleaning tools You may further clean the gold cathode with ultrasonic treatment at low power for up to 30 seconds while immersing the tip of the sensor head in distilled water In extreme and rare c...

Page 4: ...rystals Fig 4 As electrolyte use KCl solution 1 mol dm 3 74 56 g potassium chloride per litre in distilled water Add distilled water to the electrolyte powder Fig 1 up to the 10 ml mark Alternatively dissolve 1 49 g KCl in distilled water with a total volume of 20 ml Store at room temperature or 4 C in a closed bottle To prevent the formation of gas bubbles in the electrolyte heat the electrolyte ...

Page 5: ...2k is switched on Such stabilization may require several hours For this purpose the O2k may be left on overnight 4 Electrical cable connection Connect the POS cable to the O2k Main Unit avoiding bending and torsions of the cable Insert the male plug of the POS cable into the female connector next to the control light of the stirrer The red dot on the male plug has to face accurately upwards when i...

Page 6: ...t or drying the sensor body at 60 C for a few days the zero current should be reduced In addition the zero current of the bare cable without sensor head connected is tested for any leak currents If the latter test excludes any sources of leak currents other than the POS and POS service is not successful the sensor head must be replaced A new sensor head can be screwed onto the sensor connector if ...

Page 7: ...ights of the Oxygraph 2k left chamber between the MAINS and STIRRER lights right chamber between the STIRRER and COMM lights The ridge on the male plug has to face accurately downwards when inserting the plug before rotating the connector screw and fixing it finger tight O2k Manual MiPNet19 01A Oxygraph 2k start high resolution respirometry MiPNet14 01 Electrostatic discharge ESD damage and protec...