The following treatment values can 
be stored by means of the hand-held
programming unit. 
(7)

– Knee extension

– Knee flexion

– Pause extension

– Pause flexion

– Force

– Speed

IMPORTANT:

It is possible to program single or all
parameters. If only some parametres
were changed, the other parameters 
will be saved with current settings.

1. Pressing the Extension and STOP

keys at the same time for one 
second or holding down the STOP
key for five seconds enables you to
change to programming mode.

11

10

Warm up allows the patient gradually
attain full programmed range of motion.
The device starts in the middle between
the two values set for extension and
flexion. With each movement cycle the
extent of movement is increased by 
2 degrees until the set value is reached.
The device then moves between these
values.

The full speed & motion function is only
for service. The device runs at twice 
the maximum programmable speed to
facilitate a rapid device set up.

WARNING: Do not run the device in
full speed & motion when patient is 
in the device!

The individual run time for each treat-
ment. To reset press SET key in the
programming mode.

The total device run time is counted
from the first usage of the device. Press
button for 5 seconds until setting
appears. Device run time cannot be
deleted.

To save the programmed special func-
tions, press the STOP key.

Press the START key: the device checks
programmed values.

Adjusting the force (reverse on load)

– Minimum setting for reverse on load:

25 kp

– Maximum setting for reverse on load:

45 kp

Settings are aproximate!

Tensile force is measured on the frame
around the foot.

The input setting determines the maxi-
mum resistance needed to automatical-
ly reverse the direction of motion.

Speed

Minimum setting for speed: 1%

Maximum setting for speed: 100 %

Special functions are: 

– Center warm up

– Full speed & motion (double speed

setting)

– Runtime (patient runtime)

– Device runtime

Programming the special functions:

1. Switch to programming mode 

(section 5.1.1)

2. Press FUNC key

3. Select special funtions using 

or key

4. Follow the instructions on the display

5. Quit and save with STOP button

2. You can now select the treatment 

values in succesion by pressing the
parameter keys.

3. Change the value by pressing the 

+/- keys.

4. Continue programming (with 2 and 3)

until all required values are entered.

5. Press the STOP key to save all 

previous values.

6. Press START button: programme 

values were checked automatically.

7. Press START button again to start 

the device in therapy mode.

8. Pressing the parameter buttons 

in stop mode the display shows the
current stored values.

Setting the range of motion ROM

– Maximun knee extension: -10 degrees

– Maximum knee flexion: 125 degrees

The criterion for correct adjustment is
that it should be possible to move the
extremity without pain or irritation.

Adjusting the pauses

– The pauses occur in the final position

of extension of flexion and can be set
seperately for extension and flexion.

– Possible values for pauses: 

0–30 seconds.

Fig. 1

Buttons to 
choose the 
parameters:

Extension

Flexion

Pause 

Extension

Pause 

Flexion

Force

Speed

Display

STOP

START

“SET“

“+” button

“–” button

Function

5. Setting the treatment values

5.1 Programming the

ARTROMOT

®

-K4

5.1.1 Programming the

treatment values

5.1.2 Information about

treatment values

PRECAUTION!

The programmed value and 
the actual angle measured at the
patient´s knee may vary.

Center warm up

Full speed & motion

Run time

Device run time

Save data

PRECAUTION!

The reverse circuit is purely a 
safty measure for cramps, spasms,
locked joints, etc. The manufac-
turer accepts no liability if used
improperly.

5.1.3 Programming the

special functions

Summary of Contents for ARTROMOT-K4

Page 1: ...Operation Manual ARTROMOT K4...

Page 2: ...2 12 13 17 11 10 9 14 8 1 2 5 15 16 3 4 7 6 18 19 19 20 21 22 23 3 Device description Figures Fig 2 Fig 3 Fig 4 Fig 5 Fig 6 Fig 7 Fig 8...

Page 3: ...hroscopy procedures in combination with synovectomy arthrolysis or other intra articular measures mobilsations of joints in narcosis operative treatments on fractures pseudarthrosis and inversion oper...

Page 4: ...persons as otherwise all warranty services and liabilities shall be void Perform regular checks on all compo nents for possible damage or loose connections Damaged or worn parts should be replaced im...

Page 5: ...OMOT K4 features a true anatomical knee and hip axis for maximum patient comfort ARTROMOT K4 has to be set up either for the right or left leg The device can be converted quickly The procedure is easi...

Page 6: ...Speed Minimum setting for speed 1 Maximum setting for speed 100 Special functions are Center warm up Full speed motion double speed setting Runtime patient runtime Device runtime Programming the spec...

Page 7: ...lative humidity 20 to 85 Air pressure 700 hPa to 1060 hPa Operational Conditions Surrounding temperature 50O F to 104O F Relative humidity 30 to 75 Air pressure 700 hPa to 1060 hPa 13 12 6 Maintenance...

Page 8: ...dical Devices Part 1 and 2 additional norm electromagnetic compatibility requirements and testing The adherence to the standard specifications entitles to marking of these devices with CE 0297 0297 Fr...

Reviews: