orlvision RSX-P Instructions For Use Manual Download Page 7

 

  the Video Rhino Laryngoscope handpiece with insertion tube and  

the connector for the USB connection and the  

  Connection Cable RSX-USB / RSX-P (Article no. M-820-02001-0046) 

  A hand air pump (Leak Tester; item no. M-860-00003-0057).  

for carrying out the leak test. 

  The OrlView software 

  These instructions for use or a note on downloading from the homepage 

  Cover cap " Plug Cap Naso " for 9-pole plug connection 

  Cover cap for connection pressure test, article no. M-860-0003-0086-P 

 

5  Technical data 

Parameter 

Data RSX-USB 

Data RSX-P 

Focus area 

10 - 55 mm 

10 - 55 mm 

Angle of view 

90° 

90° 

Diameter distal end  

3.9 mm 

2.9 mm 

Diameter insertion tube 

3.9 mm 

2.9 mm 

Distal angulation up / down 

120° / 120° 

120° / 120° 

Working length 

310 mm 

310 mm 

Total length 

540 mm 

540 mm 

Resolution in pixels 

400x400 

400x400 

White balance 

Factory set 

Factory set 

Lighting: LED internal 

Fibre optic cable 

Fibre optic cable 

Power supply  

5V DC / 0.3 A via USB

 

5V DC / 0.3 A via USB

 

Power  

max. 2.2 W 

max. 2.2 W 

Interface 

USB 2.0 "A 

USB 2.0 "A 

Weight in g 

approx. 320g handpiece 

approx. 320g handpiece 

Risk class according to MDR 

Transport and storage temperature in ° Cel-
sius 

- 10°C to + 60°C 

- 10°C to + 60°C 

Operating temperature in ° Celsius 
The distal end can warm up to 9°C above 
room temperature. 

0° to + 40° 

0° to + 40° 

Relative humidity 

0 to 100% 

0 to 100% 

Air pressure 

950 to 1050 hPa 

950 to 1050 hPa 

Protection class against environmental influ-
ences 

IP 68 

IP 68 

Operating mode 

Continuous operation 

Continuous operation 

Table1 
 
 

 

Summary of Contents for RSX-P

Page 1: ...Rhino Laryngoscope Type RSX USB RSX P Picture 1 Video Rhino Laryngoscope 1 Integrated LED 4Status LED 2 USB interface 5 Release button 2 1 Tablet PC 6 Probe control 2 2 USB 7 Distal end 3 Power on off...

Page 2: ...insertion tube 9 7 3 Switch off 9 8 Cleaning care and disinfection 10 8 1 Manual procedures 10 8 1 1 Cleaning 10 8 1 2 Disinfection 10 8 1 3 Final rinse 11 8 2 Machine processes 11 8 3 Leak test 12 8...

Page 3: ...scope is a high quality precision mechanical optical instrument handle it with care Caution Do not use the Video Rhino Laryngoscope if it has any defects that could endanger patients or third parties...

Page 4: ...e sure that no stored image is being viewed when current examinations are to be carried out are to be performed 5 Avoid direct sunlight X rays sudden sharp temperature changes or heating above 60 C an...

Page 5: ...s used exclusively in medical practices and clinics by staff specially trained in handling en doscopes Attention The Video Rhino Laryngoscope RSX USB RSX P may only be used on persons who have a suffi...

Page 6: ...e observation region is obtained from an LED integrated in the handpiece The distal end of the insertion tube can be angled in a range of 120 by operating the angulation lever Power is supplied via th...

Page 7: ...mm 2 9 mm Distal angulation up down 120 120 120 120 Working length 310 mm 310 mm Total length 540 mm 540 mm Resolution in pixels 400x400 400x400 White balance Factory set Factory set Lighting LED int...

Page 8: ...cturer The manufacturer of the Video Rhino Laryngoscope RSX USB RSX P is orlvision GmbH Industrial Road 17 D 35633 Lahnau Tel 49 0 64 41 67 92 98 0 Fax 49 0 64 41 67 92 98 99 info orlvision de www orl...

Page 9: ...e insertion tube into the corresponding regions to be examined nasal and pharyn geal cavities If necessary the distal end can be adjusted 120 with the adjustment lever see fig ure 3 The plane of the a...

Page 10: ...ata sheet available at www wetwipe eu Attention Please do not exert any great mechanical forces on the flexible end of the endoscope when wiping it endoscope the light guides inside could be damaged b...

Page 11: ...ctor from CANTEL GmbH or an equivalent machine with the settings given below Cleaning Cleaning agent 0 5 Dr Weigert neodisher Mediclean forte Automatic cleaning process with programme no 24 with the f...

Page 12: ...at 20 C for 4 minutes Attention Usually the laryngoscope is connected to the automatic pressure monitoring during machine cleaning If this is not done we recommend using the optionally available cover...

Page 13: ...erwise moisture may enter the unit and repair may be necessary After completing the leak test open the valve on the leak tester to release the excess pressure Then turn the tester connection cap to th...

Page 14: ...mation on the performance characteristics Key features of the Video Rhino Laryngoscope are Display of images of the examination re gion the nose pharynx and larynx nasal cavities and nasopharynx If th...

Page 15: ...adiated Emission according to CISPR 11 level according to EN 55011 30 MHz to 1 GHz Class B Conducted emission according to CISPR 11 level according to EN 55011 with 150 kHz 30 MHz Class B 12 Reporting...

Page 16: ..._______________________________________________________________ _________________________________________________________________________________ ______________________________________________________...

Page 17: ...17 orlvision GmbH Gewerbestrasse 17 D 35633 Lahnau Tel 49 0 6441679298 0 Fax 49 0 6441679298 99 info orlvision com www orlvision com Rev 21 from 01 06 2022...

Reviews: