![orlvision RSX-P Instructions For Use Manual Download Page 7](http://html1.mh-extra.com/html/orlvision/rsx-p/rsx-p_instructions-for-use-manual_1653904007.webp)
7
the Video Rhino Laryngoscope handpiece with insertion tube and
the connector for the USB connection and the
Connection Cable RSX-USB / RSX-P (Article no. M-820-02001-0046)
A hand air pump (Leak Tester; item no. M-860-00003-0057).
for carrying out the leak test.
The OrlView software
These instructions for use or a note on downloading from the homepage
Cover cap " Plug Cap Naso " for 9-pole plug connection
Cover cap for connection pressure test, article no. M-860-0003-0086-P
5 Technical data
Parameter
Data RSX-USB
Data RSX-P
Focus area
10 - 55 mm
10 - 55 mm
Angle of view
90°
90°
Diameter distal end
3.9 mm
2.9 mm
Diameter insertion tube
3.9 mm
2.9 mm
Distal angulation up / down
120° / 120°
120° / 120°
Working length
310 mm
310 mm
Total length
540 mm
540 mm
Resolution in pixels
400x400
400x400
White balance
Factory set
Factory set
Lighting: LED internal
Fibre optic cable
Fibre optic cable
Power supply
5V DC / 0.3 A via USB
5V DC / 0.3 A via USB
Power
max. 2.2 W
max. 2.2 W
Interface
USB 2.0 "A
USB 2.0 "A
Weight in g
approx. 320g handpiece
approx. 320g handpiece
Risk class according to MDR
1
1
Transport and storage temperature in ° Cel-
sius
- 10°C to + 60°C
- 10°C to + 60°C
Operating temperature in ° Celsius
The distal end can warm up to 9°C above
room temperature.
0° to + 40°
0° to + 40°
Relative humidity
0 to 100%
0 to 100%
Air pressure
950 to 1050 hPa
950 to 1050 hPa
Protection class against environmental influ-
ences
IP 68
IP 68
Operating mode
Continuous operation
Continuous operation
Table1