4 Functionality and application
4.1
Application
The flexible Video Rhino Laryngoscopes RSX-USB / RSX-P from
orlvision
GmbH (hereinafter re-
ferred to as orlvision) are high-quality medical products. They are used for endoscopic examinations in
human ear, nose and throat medicine. The flexible Video Rhino Laryngoscopes can be used to exam-
ine the nasopharynx. The image of the examination region can be displayed on a PC equipped with a
high-resolution screen via the USB interface.
The RSX-USB laryngoscope is intended for use in clinics and doctors' surgeries and may only be used
by personnel specially trained in handling endoscopes!
Proceed with cleaning and disinfection as described in these instructions for use.
4.2
Image processing
Image processing software is required to display and evaluate the images. The OrlView software is
included in the scope of delivery. This is suitable for displaying the images on the optionally available
panel PC.
4.3
How it works
At the distal end of the Video Rhino Laryngoscope is the exit of a light guide that illuminates the obser-
vation region. The lens allows imaging at a 90° angle of view. The image thus captured is recorded by
a video camera, converted into an electrical signal and made available to a PC via a USB interface.
The light for illuminating the observation region is obtained from an LED integrated in the handpiece.
The distal end of the insertion tube can be angled in a range of +/- 120° by operating the angulation
lever.
Power is supplied via the PC through the USB interface.
Using the optional software, pictures or videos can be displayed and also saved.
4.4
Notes on use
The instructions for use explain how to operate the medical device safely, properly and effectively.
Please read the instructions for use before putting the device into operation, starting with the chapter
on risks and hazard warnings. Keep the instructions close to the device. Observe the ambient condi-
tions specified in the technical data.
The instructions for use do not replace the corresponding basic medical and technical knowledge. The
user may have to acquire such knowledge in special advanced training courses.
orlvision
accepts no liability for diagnoses and interpretations of findings made with the aid of medical
products acquired from
orlvision. The
acquisition of medical expertise and its diagnostic and thera-
peutic consequences are solely the responsibility of the user of the medical product.
Before each use, test the direction of movement of the angulation unit by operating the adjustment
lever to avoid an incorrect angulation direction.
We recommend the use of a lubricant on the shaft before inserting the shaft into the nasal cavity to be
examined.
4.5
Scope of delivery
The scope of delivery for the flexible Video Rhino Laryngoscope is as follows: