Caution: 

Never connect or disconnect the tester under water, otherwise moisture may enter the unit and repair 
may be necessary. 
 
After completing the leak test, open the valve on the leak tester to release the excess pressure. Then 
turn the tester connection cap to the left and then pull off the tester connection. To protect the 9-pin 
connector, we recommend using the optionally available cover cap. 
 
After completing the leak test, open the valve on the leak tester to release the excess pressure. Then 
turn the tester connection cap to the left and pull off the tester connection.  

10.4 

Care  

The Video Rhino Laryngoscope is easy to care for. Apart from thorough cleaning and regular inspec-
tion for damage, no special care is required. The laryngoscope should be stored in a dry place, safe 
from dust. 

11  Maintenance and repairs  

The components of the Video Rhino Laryngoscope are maintenance-free for their users. Repairs and 
maintenance work may only be carried out by the company 

orlvision 

or by specialist companies au-

thorised by it. The company orlvision provides the authorised companies with all necessary product 
documentation. 
 

 

Caution: 

Unauthorised opening, repairs and modifications to the laryngoscope release 

orlvision 

from any liabil-

ity for operational safety. This will void any warranty claims during the warranty period. 

11.1 

Return  

In order to avoid damage due to transport and shipping in the event of a return, please use only the 
original shipping packaging. 
 

 

Caution: 

For shipping and transport, always connect the pressure tester to the light connection with the valve 
open! 
 

12  Disposal  

Environmentally friendly disposal according to EU Directive 2012/19/EU. The appliance 
contains electronic components. To prevent environmental risks or hazards due to im-
proper disposal, the product, including accessories, must be disposed of in accordance 
with the applicable EU directives 2012/19/EU. Disposal can be carried out by the manu-
facturer. 

For this purpose, please send to the manufacturer at:  
Orlvision GmbH, Gewerbestraße 17, D-35633 Lahnau. 
Disposal in household waste is prohibited. 

13  Reporting of serious incidents  

All serious incidents related to this product shall be reported to the manufacturer and to the competent 
authority of the Member State in which the user and/or the patient is established. 
 

 

 

Summary of Contents for RS1-NTSC

Page 1: ...1 Instructions for use Video Rhino Laryngoscope RX1 RS1 PAL RS1 NTSC...

Page 2: ...ecommended accessories 9 7 1 Cold light source 9 7 2 Monitor 9 8 Connection of the Video Rhino Laryngoscope 9 9 Use of the Video Rhino Laryngoscope 10 9 1 White balance 10 9 2 Insert angle and retract...

Page 3: ...r and in proper condition including the power supply The Video Rhino Laryngoscope is a high quality precision mechanical optical instrument treat it with care Caution Do not use the Video Rhino Laryng...

Page 4: ...ge and practical experience can guarantee proper handling The operational safety and usability of the medical device depend not only on your skill but also on the care of the device Regular cleaning a...

Page 5: ...tructions for use Symbol for separate collection of electrical and electronic equipment Applied part type BF In the instructions for use Caution general danger area IP 68 Unit is dust tight and protec...

Page 6: ...ge to verify colour fidelity We recommend the use of a lubricant on the shaft before inserting the shaft into the nasal cavity to be examined 4 3 Application The flexible Video Rhino Laryngoscopes fro...

Page 7: ...ter RX1 RS1 Focus area 10 55 mm Angle of view 90 Diameter distal end 3 9 mm Diameter insertion tube 3 9 mm Distal angulation up down 130 130 Working length 310 mm Total length 540 mm Video format NTSC...

Page 8: ...nded for use in hospitals without direct connection to the public supply network The RX1 laryngoscope complies with EN 60601 1 2 2007 IEC 60601 1 2 2007 modified with regard to radiation and immunity...

Page 9: ...o Rhino Laryngoscope 1 Handpiece 2 White balance 3 Capture Button 4 Adjustment lever 5 Light guide plug 6 LED display 7 Supply hose 8 Valve for pressure test 9 Multipoint connector 10 Multipoint socke...

Page 10: ...hino Laryngoscope 9 1 White balance Attention Before each examination a white balance should be carried out so that the camera repro duces the natural colours Handpiece 1 Handpiece 2 White balance 3 C...

Page 11: ...disposable cloth afterwards Alternatively we recommend Triamin disinfectant wipes from Wet Wipe available from Orlvision Contact time depending on possible organisms 5 15 minutes see separate data sh...

Page 12: ...ries from the disinfectant solution with fresh disposable gloves Place the disinfected laryngoscope in a basin tub with microbiologically safe water drinking water quality Use fresh water for each ins...

Page 13: ...e Disinfection Disinfectant 1 0 Dr Weigert neodisher endo SEPT PAC Automatic cleaning and disinfection process with programme no 22 NORMAL PAA with the following settings Step Pre cleaning for 4 minut...

Page 14: ...d to the laryngoscope and cannot be pulled off Close the valve on the leak tester Create a test pressure of 160 10 mmHg by pumping the leak tester Figure 7 Connection pressure tester Caution If the ma...

Page 15: ...provides the authorised companies with all necessary product documentation Caution Unauthorised opening repairs and modifications to the laryngoscope release orlvision from any liabil ity for operatio...

Page 16: ...These instructions for use have been machine translated Orlvision GmbH Gewerbestrasse 17 D 35633 Lahnau Tel 49 0 6441679298 0 Fax 49 0 6441679298 99 info orlvision com www orlvision com Rev 25 25 11 2...

Reviews: