orlvision RS1-NTSC Instructions For Use Manual Download Page 11

 

11 

9.3 

Switch off  

After use, the laryngoscope must be taken out of service by disconnecting it from the supply, the cold 
light source must be switched off and then the necessary cleaning and disinfection work must be car-
ried out. See chapter 8. 

10  Cleaning, care and disinfection  

In accordance with KRINKO/BfArM recommendations, the machine method is always the preferred 
method for reprocessing. 

10.1 

Manual procedures  

10.1.1.  Cleaning  
After each use, the Video Rhino Laryngoscope must be carefully cleaned. To do this, wipe it with a 
clean disposable cloth soaked in an appropriate disinfectant on all accessible external parts.  
 
We recommend for cleaning: 2% Sekusept® active (manufacturer Ecolab). Leave on for 5 minutes. 
Make sure that the surfaces remain moist. Wipe with a dry disposable cloth afterwards. 
Alternatively, we recommend Triamin disinfectant wipes from Wet Wipe (available from Orlvision). 
Contact time depending on possible organisms 5-15 minutes, see separate data sheet, available at 
www.wetwipe.eu 

.  

 

 

Caution: 

Please do not exert any great mechanical forces on the flexible end of the endoscope when wiping it.  
endoscope, the light guides inside could be damaged (breakage).  
 

10.1.2. 

Disinfection  

 
Disinfection may only be carried out by trained personnel and in accordance with the specifications of 
the Robert Koch Institute. 
 

 

Caution: 

Before each disinfection / insertion, a leak test (see chapter 9.3) must be carried out. In case of  
the Rhino Laryngoscope must be discarded immediately and sent to the manufacturer for repair.  
manufacturer for repair. In case of leaks, the immersion disinfection becomes ineffective! 
 
We recommend the following disinfection: Immersion disinfection with 2% Sekusept® active.  
(manufacturer Ecolab). Immersion time 30 minutes.  
 

 

Summary of Contents for RS1-NTSC

Page 1: ...1 Instructions for use Video Rhino Laryngoscope RX1 RS1 PAL RS1 NTSC...

Page 2: ...ecommended accessories 9 7 1 Cold light source 9 7 2 Monitor 9 8 Connection of the Video Rhino Laryngoscope 9 9 Use of the Video Rhino Laryngoscope 10 9 1 White balance 10 9 2 Insert angle and retract...

Page 3: ...r and in proper condition including the power supply The Video Rhino Laryngoscope is a high quality precision mechanical optical instrument treat it with care Caution Do not use the Video Rhino Laryng...

Page 4: ...ge and practical experience can guarantee proper handling The operational safety and usability of the medical device depend not only on your skill but also on the care of the device Regular cleaning a...

Page 5: ...tructions for use Symbol for separate collection of electrical and electronic equipment Applied part type BF In the instructions for use Caution general danger area IP 68 Unit is dust tight and protec...

Page 6: ...ge to verify colour fidelity We recommend the use of a lubricant on the shaft before inserting the shaft into the nasal cavity to be examined 4 3 Application The flexible Video Rhino Laryngoscopes fro...

Page 7: ...ter RX1 RS1 Focus area 10 55 mm Angle of view 90 Diameter distal end 3 9 mm Diameter insertion tube 3 9 mm Distal angulation up down 130 130 Working length 310 mm Total length 540 mm Video format NTSC...

Page 8: ...nded for use in hospitals without direct connection to the public supply network The RX1 laryngoscope complies with EN 60601 1 2 2007 IEC 60601 1 2 2007 modified with regard to radiation and immunity...

Page 9: ...o Rhino Laryngoscope 1 Handpiece 2 White balance 3 Capture Button 4 Adjustment lever 5 Light guide plug 6 LED display 7 Supply hose 8 Valve for pressure test 9 Multipoint connector 10 Multipoint socke...

Page 10: ...hino Laryngoscope 9 1 White balance Attention Before each examination a white balance should be carried out so that the camera repro duces the natural colours Handpiece 1 Handpiece 2 White balance 3 C...

Page 11: ...disposable cloth afterwards Alternatively we recommend Triamin disinfectant wipes from Wet Wipe available from Orlvision Contact time depending on possible organisms 5 15 minutes see separate data sh...

Page 12: ...ries from the disinfectant solution with fresh disposable gloves Place the disinfected laryngoscope in a basin tub with microbiologically safe water drinking water quality Use fresh water for each ins...

Page 13: ...e Disinfection Disinfectant 1 0 Dr Weigert neodisher endo SEPT PAC Automatic cleaning and disinfection process with programme no 22 NORMAL PAA with the following settings Step Pre cleaning for 4 minut...

Page 14: ...d to the laryngoscope and cannot be pulled off Close the valve on the leak tester Create a test pressure of 160 10 mmHg by pumping the leak tester Figure 7 Connection pressure tester Caution If the ma...

Page 15: ...provides the authorised companies with all necessary product documentation Caution Unauthorised opening repairs and modifications to the laryngoscope release orlvision from any liabil ity for operatio...

Page 16: ...These instructions for use have been machine translated Orlvision GmbH Gewerbestrasse 17 D 35633 Lahnau Tel 49 0 6441679298 0 Fax 49 0 6441679298 99 info orlvision com www orlvision com Rev 25 25 11 2...

Reviews: