Step 5: Customise the CO2 alarm. 

 

The audible alarm is set by default to go off at 1000 ppm.

 

 

However,  in  order  to  adapt  to  different  needs  and  regulations,  the  alarm  triggering  measure  is 
customisable. This does not change the measurement system. 

 

In normal display mode, press and hold the "

-

" button (5) until the CO2 value flashes and  

the icon         appears. 

 

Press the "

+" 

and "

-

" buttons to increase or decrease the CO2 alarm threshold value. Press button (3) 

to confirm and return to the display. 

 

Caution: 

For  safety  reasons,  the  alarm  threshold  for  the  CO2  concentration  level  can  only  be  set 

between 500 ppm and 3000 ppm. 

 

To  activate  the  CO2  alarm  function: 

in  normal  display  mode  (not including  settings),  press  the  "

-" 

button briefly and the icon  

will appear to the left of the CO2 measurement.  

 

Do this again to deactivate the CO2 alarm. 

 

When the set threshold is exceeded, the alarm sounds and the icon  

  flashes. 

 

Step 6: Display of the minimum and maximum values. 

 

The Quaelis 14 meter displays the recorded minimum and maximum values for CO2, temperature and 
humidity. 

 

In normal  display  mode,  press  the  "

+" 

button  on  the  knob  (5) 

briefly → the 

MAX 

symbol  (above  the 

temperature)  is  displayed  as  well  as  all  the 

MAXI  values  →  press  "

+

→ 

MIN 

appears  (above  the 

humidity) as well as all the MINI values 

→ press "

+

" to return to normal display mode 

 

While viewing the recorded minimum and maximum values, press and hold the "

+" 

button (5) to reset. 

The display shows "

- - - -

" to confirm. 

 

Attention: 

When  displaying  the  minimum  and  maximum  values  for  CO2,  the  smiley  face  does  not 

appear. This is normal. The meter allows to save the minimum and maximum values for 24H and the 
values will be automatically deleted and reset every day at 00:00. 

 

Brightness adjustment : 

 

The Quaelis meter allows you to adjust the brightness on 3 levels: press successively on the button (2) 
to increase or decrease the brightness of the screen. 

 

Low battery indicator: 

When the battery is low the icon  

flashes 

and the display shows “

- - - -

“ 

and the display brightness is automatically reduced. 

 

Calibration function : 

 

 

Sensors can wear out over time and depending on the level of pollution: measurements and accuracy 
can drift. To avoid this problem and to ensure the accuracy of the measurements, the Quaelis 14 meter 
has the ABC (Automatic Baseline Calibration) technology which ensures a regular automatic calibration 
of the device. 

 

Caution: 

The Quaelis 14 is equipped with ABC calibration technology which automatically ensures the 

correct functioning of the device. 
If the user wishes to perform a manual calibration, it is necessary to ensure that the meter is placed in 
a 400 ppm outdoor environment to perform this process. To do this the meter must be in the ON position 
and the battery must be fully charged. When not in setup, press and hold the "

+" 

button (5), "CAL" will 

flash. Press button (3) to start the manual calibration.  

 

Calibration can take up to 30 minutes. Do not turn off the meter or press any buttons during calibration. 
When calibration is complete, the display will return to the CO2 measurement. 

 