76
BT37 IFU
5.
Specifications
5.1
System specifications
Dimensions
420 mm wide x 270 mm deep x 210 mm high.
Weight
15.5 kg
Storage temperature
-10 °C to +50 °C
Storage humidity
5% to 95% relative humidity non-condensing
Storage special instructions
Recharge every 4 months by connecting to the mains power supply
for 24 hours.
Operating environment
For indoor use only
Operating temperature
+5 °C to +40 °C for safe operation
See the table Control on page 77 for control limitations..
Operating humidity
5% to 90% relative humidity non-condensing
Altitude
Up to 2000 m
Pollution degree
Pollution degree 2 (BS EN61010-1)
IP rating
IP31
Electromagnetic environment
The equipment is intended for use in a basic electromagnetic
environment, characterized by being supplied directly at low voltage
from the public mains network.
Summary of Contents for BT37
Page 1: ...MA101414 ML BT37 IFU 0088 EN...
Page 2: ......
Page 3: ...3 I...
Page 15: ...15 Left arrow decrements selected digit Enter key moves to the right...
Page 21: ...21 II...
Page 24: ...24 BT37 IFU Blow through pipes with medical grade gas to clear any foreign bodies...
Page 25: ...25 Do not over tighten...
Page 26: ...26 BT37 IFU...
Page 27: ...27 Do not over tighten Connect up to 10 incubators in series Check for leaks...
Page 29: ...29 Remove cap Refit cap Ensure tubes aligned correctly Fit filter...
Page 35: ...35 2 3 External data collection 2 4 Alarm connection...
Page 36: ...36 BT37 IFU 2 5 Connecting to the mains supply...
Page 37: ...37 III...
Page 40: ...40 BT37 IFU 3 2 Setting the password Set passwords to prevent unauthorised access...
Page 43: ...43 3 3 1 1 1 Changing the left hand chamber temperature...
Page 47: ...47 Flow increased for purge duration after lids are shut...
Page 55: ...55 3 4 Changing the humidifier...
Page 56: ...56 BT37 IFU See Installing the humidifier on page 28...
Page 57: ...57...
Page 58: ...58 BT37 IFU 3 5 Switching off 3 6 Battery backup...
Page 59: ...59...
Page 60: ...60 BT37 IFU...
Page 61: ...61 IV...
Page 67: ...67 4 3 Safety testing Check for leaks...
Page 69: ...69 4 6 Troubleshooting...
Page 70: ...70 BT37 IFU...
Page 71: ...71...
Page 73: ...73 4 9 Disposal Do not dispose of with general waste Clean and disinfect first see page 63...
Page 74: ...74 BT37 IFU...
Page 75: ...75 V...