1.6 Safety
Operating the equipment in a manner not specified within this manual or under conditions outside of
the equipment specifications, may result in the protection offered by the equipment being impaired.
CO₂ gas is an asphyxiant.
Use in well ventilated areas.
Consider CO₂ alarms in confined spaces.
The release of gas could result in oxygen depletion: a risk assessment should be undertaken to deter-
mine whether oxygen depletion alarms should be installed.
Using default settings, the supplied gas is released to the room at the following rates.
Operating condition
Gases released.
Normal
Gas mix. 30 mL/min.
After lid closure
Gas mix. 360 mL/min for 3 minutes.
After bottle change
Gas mix. 360 mL/min for 9 minutes.
Additional ventilation may be required.
Equipment must be earthed. Class 1.
Ensure that the equipment and mains cords are regularly checked by a competent person, using a
Portable Appliance Tester or similar equipment, to ensure adequate earth bonding.
Ensure that the earth continuity of the mains installation is regularly inspected by a competent person.
Supply power via a residual current circuit breaker (RCCB) operating at a differential of 30 mA.
Check that the rating label matches the local mains supply voltage.
Mains lead to the power supply is the main disconnect device. If power needs to be disconnected im-
mediately, disconnect the mains lead from the power supply or switch off at the mains power outlet.
Ensure that the equipment is positioned so that the mains lead can be easily disconnected.
Connected devices must comply with EN60950 or its equivalent.
Any circuit connected to the alarm output must meet the requirements for an accessible part as de-
fined in EN 61010-1 or its equivalent.
Do not connect to Ethernet local area networks (LAN) external to the building.
User servicing limited to cleaning and calibration.
To avoid risk of fire, fuses must always be replaced with the same type and rating.
Fuses should only be replaced by suitably trained service personnel.
Fuses should only be replaced after the cause of the original failure has been determined and correct-
ed as appropriate.
Samples may present other biological hazards. Refer to the person responsible for the equipment.
Summary of Contents for BT37
Page 1: ...MA101414 ML BT37 IFU 0088 EN...
Page 2: ......
Page 3: ...3 I...
Page 15: ...15 Left arrow decrements selected digit Enter key moves to the right...
Page 21: ...21 II...
Page 24: ...24 BT37 IFU Blow through pipes with medical grade gas to clear any foreign bodies...
Page 25: ...25 Do not over tighten...
Page 26: ...26 BT37 IFU...
Page 27: ...27 Do not over tighten Connect up to 10 incubators in series Check for leaks...
Page 29: ...29 Remove cap Refit cap Ensure tubes aligned correctly Fit filter...
Page 35: ...35 2 3 External data collection 2 4 Alarm connection...
Page 36: ...36 BT37 IFU 2 5 Connecting to the mains supply...
Page 37: ...37 III...
Page 40: ...40 BT37 IFU 3 2 Setting the password Set passwords to prevent unauthorised access...
Page 43: ...43 3 3 1 1 1 Changing the left hand chamber temperature...
Page 47: ...47 Flow increased for purge duration after lids are shut...
Page 55: ...55 3 4 Changing the humidifier...
Page 56: ...56 BT37 IFU See Installing the humidifier on page 28...
Page 57: ...57...
Page 58: ...58 BT37 IFU 3 5 Switching off 3 6 Battery backup...
Page 59: ...59...
Page 60: ...60 BT37 IFU...
Page 61: ...61 IV...
Page 67: ...67 4 3 Safety testing Check for leaks...
Page 69: ...69 4 6 Troubleshooting...
Page 70: ...70 BT37 IFU...
Page 71: ...71...
Page 73: ...73 4 9 Disposal Do not dispose of with general waste Clean and disinfect first see page 63...
Page 74: ...74 BT37 IFU...
Page 75: ...75 V...