DP-100SL
35
Site Preparation Guide
Chapter 8
Air
Compressed Air
shows the requirements for compressed air at the site.
It is the customer’s responsibility to maintain filtering of the outside
supply to ensure an oil-free compressed air supply.
Table 8.1 Compressed Air Requirements
Pressure
0.7 ~ 1.0 MPa (7 - 10 bar)
Air supply rate
55 liters/minute
Quality
Dry (water: 0.03 g/m³)
Filtered (dust: 0.1 µm, 0.1 mg/m³)
Free of oil (oil: 0.01 mg/m³)
Tube connectors
10 mm (outside diameter)