14
7
ADDITIONAL FUNCTIONS in the MULTIMETER MODE
The key
MENU
opens the Functions
TARA
and
FILTER
for the Multimeter:
7.1
ADDITIONAL FUNCTIONS
The numeric keyboard permits selection of following functions:
HOLD
Display Hold (with CLEAR return)
MAX
Maximum display reading of the Multimeter (with CLEAR return)
MIN
Minimum display reading of the Multimeter (with CLEAR return)
CLEAR
Cancel the above functions and switches into the measuring mode
8
SOFTWARE - CALIBRATION
The voltage and the current ranges of the multimeter as well as the output signals of the calibrator can
be calibrated from the keyboard. The calibration menu is password protected. The calibration menu
opens with the key
OK
pressed for 10 seconds.
Full calibration of all ranges or partial calibration of some selected ranges can be performed. A precision
5 digits multimeter with accuracy of 0.005% is required.
Point 1:
Calibrator Outputs
Points 2-3:
Multimeter Inputs
Point 4:
Calibration of the Resistor Simulator
Point 5:
Hardware Configuration
Point 6:
Selection of Logo for customized instrument
Quit the Menu with
ESC
.
To enter the Calibration Menu
use the factory set Password “8952”. The Password can be
changed in step “6”. NOTE THE NEW PASSWORD! If lost, consult the manufacturer.
The TARE can be activated
The Filter can be set for the input signal
.