FUNCTION SELECTION - CALIBRATOR
The calibrator function is selectable with the keys. The output values can be set in steps, ramps or free
values.
The output function will be selected with
OUT
. Vertical cursor keys select the required function:
Current Source 0 - 22mA
Current Sink 0 - 22mA
Voltage Output 0 - 25V
Voltage Output 0 - 560mV
Voltage Output 0 - 28mV
Resistance Output 45 - 3000 Ohm
RTD Outputs Pt-100, Pt-200, Pt-500,
Pt-1000, Ni 1000
Simulation of Thermocouples
B, J, K, N, R, S, T
OK confirms the selection. The selected function remains memorised also when the power is switched-off.
If the resistance is measured with external Ohm-meter, the polarity of the Ohm-meter
input and the OC505 outputs has to be identical.