1* WARNING: Use of this equipment adjacent to or stacked with other equipment
should be avoided because it could result in improper operation. If such use is
necessary, this equipment and the other equipment should be observed to verify that
they are operating normally.
2* WARNING: Use of accessories, transducers, and cables other than those specified
or provided by the equipment manufacturer could result in increased electromagnetic
emissions or decreased electromagnetic immunity of this equipment and result in
improper operation.
3* WARNING: Portable RF communications equipment (including peripherals such as
antenna cables and external antennas) should be used no closer than 30 cm (12
inches) to any part of the ME equipment, including cables specified by the manufactur-
er. Otherwise, degradation of the performance of this equipment could result.
4* WARNING: You might see some standard deviation in measurements.
-- Blood Oxygen Saturation Measurement Accuracy: ±3%
-- Pulse Rate Measurement Accuracy: ± 2BPM
5* WARNING: The FUNCTIONAL TESTER in the Smart Wrist, RC178LW, cannot be
used to determine the ACCURACY of a PULSE OXIMETER PROBE or a PULSE OXIME-
TER MONITOR.
6* WARNING : This product can expose you to the chemical lead known to the State
of California to cause cancer. For more information, go to www.P65Warnings.ca.gov.
7* WARNING : Software algorithms and DATA UPDATE PERIOD may delay the dis-
played and transmitted values of SpO2 and Pulse Rate.
Table 1
declaration - electromagnetic emission
Emissions test
Compliance
RF emissions
CISPR 11
Group 1
RF emissions
CISPR 11
Class B
Harmonic emissions
IEC 61000-3-2
Not applicable
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Not applicable
27