10 2
129
11 5
130
12 2
129
13 4
130
14 2
129
15 5
130
16 1
129
17 4
130
…
First value is column, second value is Avg(DP)-UserGG(x)*DP(x) and third is the column gain
UserGG(x) (coded in decimal: floating value is decimal column gain divided by 128).
After that, click on ‘Write Param (CP80-4) to camera’ to upload the parameters to the camera.
To activate FPN&PRNU parameters, check ‘Load param on boot’ to enable the camera to load
the calibration parameters at the boot.
Wait until finished and reboot. The camera leds will ‘swipe’ in red for 10 seconds and then the
camera is ready.
Activate correction (FPN or FPN/PRNU)
By default, no correction is activated. The XML correction field is defined in Optronis-
>Correction.