Technical Data
29
-
2.5 Cooling properties
Flow
Temperature at
1 l/min
Temperature at
2,5 l/min
Temperature at
5 l/min
Ambient temperature
[255 °C]
Cooling water input [°C]
19
19
19
Camera/ Thermometer [°C]
36
34
32
Table 1:
Cooling properties with a steady cooling water input temperature and various flow
Flow
Temperature at 2.5 l/min of flow
Ambient temperature
[255 °C]
Cooling water input [°C]
25
30
35
40
Camera/ Thermometer [°C]
39
42
45
51
Table 2:
Cooling properties with a steady flow and various cooling water input temperature
2.5.1 Condensation
For applications at ambient temperatures until 100 °C and a high humidity there is danger of
condensation (see
Table 3
).
To avoid condensation, the temperature of the cooling media and the flow rate must ensure a
minimum device temperature.
Consider the operation temperature of the applied devices.
Summary of Contents for CoolingJacket Advanced
Page 18: ...18 Figure 4 CoolingJacket Advanced standard version top view...
Page 19: ...Technical Data 19 Figure 5 CoolingJacket Advanced standard version front view...
Page 20: ...20 Figure 6 CoolingJacket Advanced standard version complete view...
Page 21: ...Technical Data 21 Extended version Figure 7 CoolingJacket Advanced extended version side view...
Page 22: ...22 Figure 8 CoolingJacket Advanced extended version top view...
Page 23: ...Technical Data 23 Figure 9 CoolingJacket Advanced extended version front view...
Page 24: ...24 Figure 10 CoolingJacket Advanced extended version complete view...
Page 28: ...28 2 4 3 Cable glands on the back Figure 12 Cable glands with seal insert...
Page 34: ...34 Figure 15 Assembling of the 6 9 23 26 29 33 39 41 53 72 and 80 optics...
Page 35: ...Installation 35 Figure 16 Assembling of the 13 and 15 optics...
Page 36: ...36 Figure 17 Assembling of the 60 and 90 optics...
Page 37: ...Installation 37 3 2 2 Assembling of the front part Figure 18 Assembling of the PI 2xx...
Page 38: ...38 Figure 19 Assembling of the PI 1M 05M...
Page 39: ...Installation 39 Figure 20 Assembling of the CTlaser CSlaser CTvideo CSvideo...
Page 46: ...46 Figure 26 Back side of the CoolingJacket with locking lever 1 Locking lever...
Page 47: ...Installation 47 7 Dismount in reverse order Figure 27 Unlocking of the hinge...
Page 55: ...Installation 55 Figure 35 Notch to fix the PI Netbox top view 1 Notch...
Page 67: ......
Page 68: ...optris CJA MA E2017 08 A...