Check if the " A63.7015 (A63.7006)" device is turned on
⚫
Check if necessary, programs are executing
⚫
Check if your computer has virus.
2.3. Communication failure
Problem
The error message "Cannot find Active Scan facility" pops up
Reason
Poor USB cable connection between device and computer
Solution
Reconnect the USB cable
2.4, Vacuum Failure
Problem
A continuous beep of "BB" is emitted from the internal motherboard of the device during
the vacuuming process
Reason
⚫
The vacuum pump tube between the mechanical vacuum pump and
A63.7015/A63.7006 is not properly connected
⚫
Mechanical vacuum pump failure
⚫
Molecular pump failure
⚫
The sample compartment door is not fully closed
Solution
⚫
Check the connection status of the vacuum pump between the mechanical vacuum
pump and A63.7015/A63.7006
⚫
Close the sample compartment door tightly when vacuuming
⚫
Replacing the vacuum pump oil in the mechanical vacuum pump