A12.0910
2-2-4 Assemble the Condenser
(1) Rotate the coarse focusing knob
①
to raise stage to
the highest. (See Fig. 5)
(2) Rotate the condenser up-down knob
②
to lower the
condenser bracket to the lowest.
(3) Loosen the condenser lock-screw
③
fully.
(4) Swing out the front lens of condenser with the scale
forward. Make the lock screw
④
of condenser in
alignment with the groove
⑤
of the condenser stand.
Push the condenser into the innermost of stand.
(5) Screw down the condenser lock-screw
③
, and raise
the condenser to the highest position with the
condenser up-down knob
②
.
2-2-5 Assemble the Nosepiece
(1) Loose the lock screw
①
on the arm. (See Fig. 6)
(2) Match the dovetail interface
②
of the nosepiece
with the dovetail groove of the arm, and push it to the
bottom. Tighten the lock screw
①
.
2-2-6 Assemble the Head
(1) Loosen the head lock-screw
①
fully. (See Fig. 7)
(2) From a little right position, insert the coattail
interface on the bottom of head into the hole of middle
head with a little left inclined. Keep the two eyepiece
tubes forward, and then screw down the lock screw
①
.
2-2-7 Assemble the Objective
Rotate the coarse focusing knob to lower the stage.
Install the objectives into the nosepiece from the lowest
magnification to the highest in a clockwise direction.
(See Fig.8)
★
Search and focus the sample by low magnification
objective (4X or 10X) when operating. Then get
change to the high magnification ones according
to the observation requirements.