13
Function
Parameter Settings/Options
Hold Function
C08
0 = turn off hold function
1 = Peak hold - Grabs the highest weight
2 = Manual hold - Grabs the current weight
3 = Auto hold - Automatically holds data when stable
4 = Average hold - for animal weighing, averages the weight from a
sample of 3 seconds
5 = Auto Average hold - Average hold without the need to press the
hold key
Unit Switch
C09
0 = turn off unit switch
1 = turn on unit switch
Power Saving Mode
C10
0 = turn off power saving setting
3 = turn off display if no change within 3 minutes
5 = turn off display if no change within 5 minutes
Automatic
Power Off
C11
0 = turn off auto power off
1 = power off automatically if no change within 10 minutes
2 = power off automatically if no change within 20 minutes
3 = power off automatically if no change within 30 minutes
4 = power off automatically if no change within 40 minutes
5 = power off automatically if no change within 50 minutes
6 = power off automatically if no change within 60 minutes
Alarm Setting
C12
0 = turn off alarm
1 = turn on alarm
Upper Limit Alarm
C13
Set upper limit within the max. capacity
Lower Limit Alarm
C14
Set lower limit within the max. capacity
Inner Code Display
C15
check the inner code (raw data)
Preserved menu
C16
Preserved menu
C17
Communication
Setting for the RS232
Port
C18
Set the serial interface data output method:
0 = Turn off serial interface data output
1 = Command request mode, connect computer.
2 = Print mode, connect printer
3 = Continuous sending mode, connect computer or display
Preserved menu
C19
Manual
Zero Range
C20
0 = turn off manually zero setting
1 = ±1% max capacity
2 = ±2% max capacity
4 = ±4% max capacity
Initial Zero Range
C21
0 = no initial zero setting
1 = ±1% max capacity
2 = ±2% max capacity
5 = ±5% max capacity
10 = ±10% max capacity
20 = ±20% max capacity