Hold

There are five different hold functions, you can choose the best option for you and set it 
the parameter settings 

1. Peak Hold:

 Grabs the highest weight 

(for materials testing, ie. tension and pulling 

force)

 

● Press the HOLD key then add weight to the scale

 

● The indicator will show the highest weight it recorded and hold it on the screen until 

a higher weight is placed on the scale

2. Manual Hold:

 Grabs the current weight and holds it so it will not change/fluctuate 

 

● While weighing, press HOLD and the indicator will hold the current weight on the 

screen until HOLD is pressed again

3. Auto Hold:

 If the weight on the scale is above 20d and is stable, the indicator will 

hold that weight on the screen for 3 seconds then go back to general weighing 

 

● Pressing the hold key is unnecessary, holding is done automatically when the scale 

is stable

4. Average Hold: 

Used for animal weighing, the indicator will display the average 

weight sampled from 1-9 seconds (Set in the 

HAt

 parameter setting)

 

● Add animal to scale and press HOLD

 

● Indicator screen will show “

LOC

” for 1-9 seconds, then display the average weight 

from those seconds

 

● Press HOLD again to exit holding mode

5. Auto Average Hold: 

Used for animal weighing, the indicator will display the aver-

age weight sampled from 1-9 seconds without the need to press the HOLD key.

If the weight on the scale is above 20d (scales division x 20) and is stable, the indica-

tor will start grabbing the average weight sampled from 1-9 seconds (Set in the HAt: 

parameter setting). In parameter 

Hadt

 you can set a determined loading time from 

1-9 seconds. This is the time you have to load the animal before the scale begins to 

take the average weight. 

 

● Add animal to scale. Once weight is added to the scale, the scale will wait the set 

loading time and then the Indicator screen will show “

LOC

” for 1-9 seconds

 

● It will then display the average weight from those 1-9 seconds for 3 seconds and 

then repeat the process

Summary of Contents for OP-902

Page 1: ...optimascale com OP 902 INDICATOR USER S MANUAL...

Page 2: ...1 Features 2 Technical Parameters 2 Power Supply 2 Display and Key Descriptions 3 Operating Instructions 4 Calibration 5 6 Indicator Parameter Settings 7 Helpful Definitions 8 Troubleshooting 8 Conne...

Page 3: ...as it could damage the load cell inside Material that has a static electric charge could influence the weighing Discharge the static electricity of the samples if possible Another solution to the pro...

Page 4: ...l zero range 10 max capacity Manual zero range 2 max capacity Resolution Display Up to 1 15 000 Display Resolution Operation temperature 10 C 40 C Operation humidity 90 RH Storage temperature 40 C 70...

Page 5: ...d Confirm ZERO Zero s the scale Tare s the scale Move Left UNITS Shifts between weighing units kg lb Move Down DISPLAY Meaning The scale is at zero The scale is stable Gross Shows you are in Gross wei...

Page 6: ...ot at gross zero due to material build up If the weight on the scale is within the zero range 2 of the max capacity and stable and ZERO is pressed the indicator will zero the scale and stay in gross w...

Page 7: ...nimal weighing the indicator will display the average weight sampled from 1 9 seconds Set in the HAt parameter setting Add animal to scale and press HOLD Indicator screen will show LOC for 1 9 seconds...

Page 8: ...low for other variances Press UNIT to change the capacity range to your scales capacity Calibration chart for kilogram usage Display Max Capacity Division Calibration Loading Weight FS 0 15 00kg 0 005...

Page 9: ...ift and UNIT to change the calibration loading weight Then press HOLD to confirm The display will show CALon please follow the steps for FS values of 0 6 to complete calibration For FS values of 0 6 P...

Page 10: ...ample of 3 seconds 5 Auto Average hold Average hold without pressing HOLD Hold Time Ht 1 0 turn off hold time 1 15 second hold time 2 30 second hold time 3 45 second hold time 4 60 second hold time 5...

Page 11: ...9 Function Display Parameter Settings Baud Rate bAud 1 1 9600bit s 2 4800bit s 3 2400bit s 4 1200bit s 5 600bit s...

Page 12: ...erred in a communication channel example In the serial port context 9600 baud means that the serial port is capable of transferring a maximum of 9600 bits per second TROUBLESHOOTING Error Reason Solut...

Page 13: ...uick Disconnect DB9 Connection 9 pin Serial Connector The DB9 9 pin serial connector is used for different purposes depending on the indicator model Figure 4 shows the pin assignment on the DB9 9 pin...

Page 14: ...2 port and the indicator will respond with the current data Communication Format is done using ASCII as shown below CR LF S 1 S 2 S 3 Data S 4 S1 weight status ST standstill US not standstill OL overl...

Page 15: ...Please e mail sales optimascale com for any sales related questions Please e mail support optimascale com for any support related questions Don t forget to visit our website at optimascale com 2 5 2...

Reviews: