2.0 ASSEMBLING
- Before assembling for fluorescence parts, please
set power supply at “OFF” and take off the plug in
order to operator’s safety.
2-2-2 Set the fluorescence blocks
(1) Loose the lock-screw 1 with M4 inner hexagon spanner
and take off right cover 2 (Fig. 4)
(2) Match the dovetail glut of filter block B1 3 to the groove
of sliding channel 4, and push 3 into sliding channel
4 at right.
(3) Put on the right cover 2, point the screw 1 to the hole
5 on sliding channel 4 and tight it with M4 spanner.
(4) Insert the pole 7 into the hole 6 on right cover 2, and
point it to the hole at side of filter block B1 3, then tight.
(Fig.5)
(5) Loose the screw cap 9 a little, and turn the remark block
8 to make the letter “B” point to operator, then tight 9.
- XDS-3FL fluorescence microscope can only be
equipped
with 2pcs filter blocks. The standard is B1 and G1.
Others for selection.
- Use the same way to set filter block G1 to the left.
- Please be aware the remark (such as: DMB1) on the
side
should be correctly, do not overturn with up and down.
- During observation, both right and left cover can’t be
taken off because it protects the filter while filter
exchanging.
2-2-2 Set the fluorescence connector
(1) Hook the hole 1 to take off the fluorescence hole cover
2. (Fig. 6)
(2) Insert the fluorescence connector 3 to the fluorescence
hole
to make the screw 4 point to thread 5, and tight 3 screws
with M4 spanner. (Fig.7)
(3) Turn off one of handles 7 from fluorescence preventing
board 8 and insert the board 8 into connector, you will
find sound to mean it at middle position (there are 3
positions). And tight the handle 7. (Fig. 8)
2.2 Assembling steps: