Page 16
4.0 GENERAL SPECIFICATION
Typology
INVERTED RESEARCH MICROSCOPE
Description
Laboratory inverted microscope for research applications.
Dye-cast frame, with high stability and ergonomy, for transmitted light and and
reflected fluorescence observation.
Illumination
Transmitted Light:
Light source type X-LED8 with white 8W LED; light intensity control using a knob
on left side of the frame.
Color temperature: 6300K
LED average life time approx. 50.000h
Voltage: 110/240Vac, 50/60Hz, 1A ; Fuse: T500mA 250V
Max power required: 13W
Reflected Light:
Mercury burner 100W HBO, light control based on external power supply.
Bulb average life time approx. 300 hours.
Voltage: 10/240Vac, 50/60Hz, 1A ; Fuse: F8AL 250V.
Max power required: 125W
Observation Modes
Brightfield, phase contrast, Fluorescence B and G
Fluorescence B: EX 460-490, DM 500, EM 520LP;
Fluorescence G: EX 480-550, DM 570, EM 590LP;
Fluorescence UV (optional): EX 325-375, DM 400, EM 420LP;
Fluorescence V (optional): EX 385-425, DM 440, EM 455LP.
Fluorochromes
Excitation B: Acridine Yellow, Acridine Orange, Auramine, DiO, DTAF, FITC, GFP,
YFP, ecc.
Excitation G: DiL; Blu Evans, Feulgen, Rhodamine, Texas Red, TRITC, PI, ecc.
Excitation UV (optional):
AMCA, AutoFluorescence, BAO, BFP, Blu Cascade, DANS, DAPI, Hoechst, Indo-
1, SITA, ecc.
Excitation V (optional): ANS, Fluorescamine, Catecholamine, ecc.
Focusing
Coaxial coarse and fine focusing mechanism (graduated, 0.002mm) with upper
stop, to prevent the contact between objective and specimen. Adjustable tension of
coarse focusing knob.
Stage
Fixed stage, dimensions 250x230 mm.
Mechanical stage mountable on the right side of the stage, X-Y translation range
114x81 mm, with metallic interchangeable inserts for slides, Petri dishes, Terasaki,
multi-Well plates, etc.
Pair of side extensions to expand the surface of the stage.
Glass stage insert with hole for small dimension specimens.
Nosepiece
Quintuple revolving nosepiece, rotation on ball bearings.
Head
Trinocular observation head, inclined 45°.
Diopter adjustment on left eyepiece.
Interpupillary adjustment 55-75 mm.
Eyepieces
Wide field eyepieces EWF10X/22 with field number 22.
Objectives
Infinity corrected optical system IOS (Infinity Optical System).
Plan-achromatic LWD objectives infinity corrected, for thickness 1.2 mm, made by
following objectives:
-) Plan-achromatic IOS FLUO LWD 10X, N.A. 0.30, W.D. 10.0 mm
-) Plan-achromatic IOS FLUO LWD 20X, N.A. 0.45, W.D. 5.1 mm
-) Plan-achromatic IOS FLUO LWD 40X, N.A. 0.65, W.D. 2.6 mm
All objectives are treated with an anti-fungus treatment.
Summary of Contents for XDS-3FL