8. Troubleshooting
Review the information in the table below to solve operating problems.
PROBLEM
CAUSE
SOLUTION
When using 10x to 100x objectives,
the field diaphragm image does not
appear.
The slide is too thick
Use slides with a thickness not ex
-
ceeding 1.2 mm
Top lens is not in
Put the top lens in the light path
In bright field the image is too bright
and with low resolution
Aperture diaphragm too closed
Open aperture diaphragm in a proper
way
Top lens is not in
Put the top lens in the light path
Condenser phase ring does not over
-
lap to the objective phase ring
An incorrect insert is inserted in the op
-
tical path
Rotate the condenser turret to insert
the appropriate optical insert for your
objective
The objective does not match the
phase ring in use
Use the objective corresponding to the
phase ring
Contrast in darkfield is poor
Top lens is not in
Put the top lens in the light path
Aperture diaphragm is closed
Fully open the aperture diaphragm
An incorrect insert is inserted in the op
-
tical path
Rotate the condenser turret to insert
the appropriate optical insert for your
objective
You’re using an inappropriate objective See paragraph 7. “Observation modes
and usable objectives”.
Darkfield ring not well centered
Center the ring correctly
Image quality when observing in polar
-
ized light is not high
Polarizer is not inserted in the optical
path
Insert the polarizer in the optical path
Analyzer is not inserted in the optical
path
Insert the analyzer in the optical path
One insert is present in the optical path Remove the insert from the optical path
While observing in DIC you can not
see the typical interference colors
Polarizer is not inserted in the optical
path
Insert the polarizer in the optical path
Analyzer is not inserted in the optical
path
Insert the analyzer in the optical path
DIC prism is not inserted in the optical
path
Insert the DIC prism in the optical path
DIC slider is not inserted in the optical
path
Insert the DIC slider in the optical path
Polarizer and analyzer are not in the
total extinction position (crossed Nicol)
Reposition the polarizer
While observing in DIC the typical in
-
terference colors are not homogene
-
ous
T
he condenser is not well centered or
is at an incorrect height
Adjust the condenser according to
Koehler’s setting
An incorrect insert is inserted in the op
-
tical path
Rotate the condenser turret to insert
the appropriate optical insert for your
objective
You’re using an inappropriate objective See paragraph 7. “Observation modes
and usable objectives”.
Summary of Contents for M-1157
Page 1: ...Model M 1157 ACCESSORIES Series INSTRUCTION MANUAL Ver 1 0 2019...
Page 19: ...Modello M 1157 Serie ACCESSORI MANUALE DI ISTRUZIONI Ver 1 0 2019...
Page 37: ...Modelo M 1157 Serie ACCESORIOS MANUAL DE INSTRUCCIONES Ver 1 0 2019...
Page 55: ...Mod le M 1157 S rie ACCESSOIRES MANUEL D UTILISATION Ver 1 0 2019...
Page 73: ...Modell M 1157 ZUBEH R Serie BEDIENUNGSANLEITUNG Ver 1 0 2019...
Page 91: ...Modelo M 1157 ACESS RIOS S rie MANUAL DE INSTRU ES Ver 1 0 2019...