or power socket.
After you turn on the power, you will hear worm gear is rotating. That means the instrument is making
self-checking and resetting. After finishing self-checking, the instrument will stop automatically and wait
for communication with computer.
Now you can run the 7ISWS application software and select the actual RS232 port to make
communication. During the self-checking, the instrument cannot communicate with computer and the
software will prompt the communication was failed. Please exit the software, after self-checking is
finished (the sound of rotation is stopped), rerun the software and select the actual RS232 port, the
communication will be successful.
You can use software to operate the instrument to automatically change the wavelength, scan spectrum,
switch gratings and change filters (if you have bought 7IFW6 filter wheel from us). Please refer to the
software manual.
The spectrometer should be power on when using CCD. The wavelength range of CCD need to be set by
the spectrometer. Please connect CCD by USB line and switch to CCD by hand or by software. Please
refer to software manual.
Note: To avoid damage of computer and instrument, please do not plug / unplug cables when the
instrument is still power on.
9.
Others
Fiber adapter
If you have bought the adapter for SMA905 fiber from us, please replace the cover of the slits with the
fiber adapter and connect the fiber directly.