7.
Slits Adjustment
This series of spectrometers uses micrometer adjustable slits with sharp stainless-steel blades. The
narrowest practically achievable width is 10 μm. The slit is 5 mm high and its width can be adjusted from
10 μm to 3 mm.
The slit uses micrometer head with 10μm graduations to drive the width of slits. The displacement is
0.25mm per revolution.
8.
Usage
Cable Connection:
Connect 9-pin connector of spectrometer to the RS232 serial port of computer. Or connect the
spectrometer to computer by USB line.
If you have bought our 7IFW6 filter wheel, please connect 15-pin connector of spectrometer to
the 9-pin connector of 7IFW6 filter wheel.
Connect the CCD to computer by USB line.
After connecting all the cables firmly, turn on the power of the instrument. The red indicator light will
be lighted up to represent normal power supply. If the red light is not lighted up, please check the fuse