4
/
6
through the BNC input plug.
d)
HV Adj: It’s used to control output voltage manually. The output
voltage will increase when turning clockwise, and please lock it after
adjusting to avoid accidental change of high voltage output.
e)
(0V to 5V) Remote In: BNC plug for external adjustment of input
voltage.
f)
HV Output: Output voltage BNC plug is used to connected to
photomultiplier tube.
2)
Operation method:
a)
Connect HV Output to photomultiplier tube.
b)
Counterclockwise adjust HV Adj to minimum.
c)
Turn on the power.
d)
Press HV Start to start high voltage output.
e)
Manually adjust HV Adj to needed voltage.
f)
For external voltage control, please connect the external control signal
to (0V~5V) Remote In and press the Control Switch.
3)
Cautions:
a)
The maximum power output is -1250V voltage, please pay attention to
the use of safety.
b)
High voltage output connection cable is high voltage resistant
equipment, please do not use wrongly.
c)
Do not expose the photomultiplier tube to strong light at high voltage
to avoid damage.
d)
Before turning on the power supply, it is best to turn the HV Adj knob
counterclockwise to make the output voltage minimum.