4-20 mA Setting
16
For the item by 4-20 mA signal output, set the measurement value to output
at lower limit (Lo, 4 mA) and upper limit (Hi, 20 mA).
20 mA
Signal output
4 mA
Measurement
value at 4 mA
Measurement
value at 20 mA
Upper/lower limit of signal output range
Output item
[Display]
Factory default
setting
Signal output range
Lower limit to Upper
limit
Minimum setting
range
Upper limit - Lower
limit
≥
Suspended
solids [SS]
Lower limit: 0.0
Upper limit: 1000
0.0 to 1000
20.0
Transparency
[Trans.]
Lower limit: 0
Upper limit: 100
0 to 100
20
The setting unit is based on the display resolution of the item set for the
display item. (For display resolution, refer to “14 Specifications” P.34)
For operation, refer to the "Universal Transmitter SC-U1 Operation
Manual", "5.3 4-20 mA Setting", "
◆
Signal output range lower limit"
and "
◆
Signal output range upper limit".