-Never modify or touch the internal structure of the device. Electric shock or malfunction may result. There are no user-
serviceable parts inside the device.
-Verify that the following specified environmental conditions for installation are met. In addition, verify that usage area
meets the following conditions.
Make certain that the instrument is operated in the following environment conditions:
Humidity 30% to 75%
No large amount of dust is contained in the air.
Subdued ambient light.
Locate instrument in a place free from vibration or impact Locate instrument on a stable surface.
Place instrument in a place free from vibration or impact Place instrument on a stable surface
Be sure that a power outlet meets the power requirements.
If the supplied voltage is too high or low, the device may not deliver full performance and malfunction, or fire may result.
Be sure to use the power supply that came with device. Using any adapter other than the supplied one may result in
malfunction, or may void the warranty.
Do not use an extension cord when supplying the device with power.
Be sure to connect using a grounded outlet. Electrical shock or fire may result in event of malfunction or electrical
leakage.
Never crush or pinch the power cord with heavy objects. Damage may result in electrical shock or fire.
Before connecting the electronic control panel cable to the instrument, turn the instrument OFF and disconnect the
power supply from the power outlet.
Usage Precautions
CAUTION
The device has been tested and found to comply with the limits for the medical devices to the IEC 60601-1-2 and
EN55011. These limits are designed to provide reasonable protection against harmful interference in a typical medical
installation. This device generates uses and can radiate radio frequency energy. If not installed and used in accordance
with instructions, may cause harmful interference to other devices in the vicinity. However, there is no guarantee that the
interference to other devices, which can be determined by turning the device off and on, is caused by this instrument.
The user is encouraged to try to correct the interference by one or more of the following methods:
Reorient or relocate the receiving device.